Combined frequency domain near-infrared spectroscopy and diffuse correlation spectroscopy system for comprehensive metabolic monitoring of inspiratory muscles during loading.

SIGNIFICANCE: Mechanical ventilation (MV) is a cornerstone technology in the intensive care unit as it assists with the delivery of oxygen in critically ill patients. The process of weaning patients from MV can be long and arduous and can lead to serious complications for many patients. Despite the known importance of inspiratory muscle function in the success of weaning, current clinical standards do not include direct monitoring of these muscles.

AIM: The goal of this project was to develop and validate a combined frequency domain near-infrared spectroscopy (FD-NIRS) and diffuse correlation spectroscopy (DCS) system for the noninvasive characterization of inspiratory muscle response to a load.

APPROACH: The system was fabricated by combining a custom digital FD-NIRS and DCS system. It was validated via liquid phantom titrations and a healthy volunteer study. The sternocleidomastoid (SCM), an accessory muscle of inspiration, was monitored during a short loading period in fourteen young, healthy volunteers. Volunteers performed two different respiratory exercises, a moderate load and a high load, which consisted of a one-minute baseline, a one-minute load, and a six-minute recovery period.

RESULTS: The system has low crosstalk between absorption, reduced scattering, and flow when tested in a set of liquid titrations. Faster dynamics were observed for changes in blood flow index (BFi), and metabolic rate of oxygen (MRO2) compared with hemoglobin + myoglobin (Hb+Mb) based parameters after the onset of loads in males. Additionally, larger percent changes in BFi, and MRO2 were observed compared with Hb+Mb parameters in both males and females. There were also sex differences in baseline values of oxygenated Hb+Mb, total Hb+Mb, and tissue saturation.

CONCLUSIONS: The dynamic characteristics of Hb+Mb concentration and blood flow were distinct during loading of the SCM, suggesting that the combination of FD-NIRS and DCS may provide a more complete picture of inspiratory muscle dynamics.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app