Near-lossless Compression of Tc-99 m DMSA Scan Images Using Discrete Cosine Transformation.
AIM AND OBJECTIVE: The objective of this study was to optimize the threshold for discrete cosine transform (DCT) coefficients for near-lossless compression of Tc-99 m Dimercaptosuccinic acid (DMSA) scan images using discrete cosine transformation.
MATERIALS AND METHODS: Two nuclear medicine (NM) Physicians after reviewing several Tc-99 m DMSA scan images provided 242 Tc-99 m DMSA scan images that had scar. These Digital imaging and communication in medicine (DICOM) images were converted in the Portable Network Graphics (PNG) format. DCT was applied on these PNG images, which resulted in DCT coefficients corresponding to each pixel of the image. Four different thresholds equal to 5, 10, 15, and 20 were applied and then inverse discrete cosine transformation was applied to get the compressed Tc-99 m DMSA scan images. Compression factor was calculated as the ratio of the number of nonzero elements after thresholding DCT coefficients to the number of nonzero elements before thresholding DCT coefficients. Two NM physicians who had provided the input images visually compared the compressed images with its input image, and categorized the compressed images as either acceptable or unacceptable. The quality of compressed images was also assessed objectively using the following eight image quality metrics: perception-based image quality evaluator, structural similarity index measure (SSIM), multiSSIM, feature similarity indexing method, blur, global contrast factor, contrast per pixel, and brightness. Pairwise Wilcoxon signed-rank sum tests were applied to find the statistically significant difference between the value of image quality metrics of the compressed images obtained at different thresholds and the value of the image quality metrics of its input images at the level of significance = 0.05.
RESULTS: At threshold 5, (1) all compressed images (242 out of 242 Tc-99 m DMSA scan images) were acceptable to both the NM Physicians, (2) Compressed image looks identical to its original image and no loss of clinical details was noticed in compressed images, (3) Up to 96.65% compression (average compression: 82.92%) was observed, and (4) Result of objective assessment supported the visual assessment. The quality of compressed images at thresholds 10, 15, and 20 was significantly better than that of input images at P < 0.0001. However, the number of unacceptable compressed images at thresholds 10, 15, and 20 was 6, 38, and 70, respectively.
CONCLUSIONS: Up to 96.65%, near-losses compression of Tc-99 m DMSA images was found using DCT by thresholding DCT coefficients at a threshold value equal to 5.
MATERIALS AND METHODS: Two nuclear medicine (NM) Physicians after reviewing several Tc-99 m DMSA scan images provided 242 Tc-99 m DMSA scan images that had scar. These Digital imaging and communication in medicine (DICOM) images were converted in the Portable Network Graphics (PNG) format. DCT was applied on these PNG images, which resulted in DCT coefficients corresponding to each pixel of the image. Four different thresholds equal to 5, 10, 15, and 20 were applied and then inverse discrete cosine transformation was applied to get the compressed Tc-99 m DMSA scan images. Compression factor was calculated as the ratio of the number of nonzero elements after thresholding DCT coefficients to the number of nonzero elements before thresholding DCT coefficients. Two NM physicians who had provided the input images visually compared the compressed images with its input image, and categorized the compressed images as either acceptable or unacceptable. The quality of compressed images was also assessed objectively using the following eight image quality metrics: perception-based image quality evaluator, structural similarity index measure (SSIM), multiSSIM, feature similarity indexing method, blur, global contrast factor, contrast per pixel, and brightness. Pairwise Wilcoxon signed-rank sum tests were applied to find the statistically significant difference between the value of image quality metrics of the compressed images obtained at different thresholds and the value of the image quality metrics of its input images at the level of significance = 0.05.
RESULTS: At threshold 5, (1) all compressed images (242 out of 242 Tc-99 m DMSA scan images) were acceptable to both the NM Physicians, (2) Compressed image looks identical to its original image and no loss of clinical details was noticed in compressed images, (3) Up to 96.65% compression (average compression: 82.92%) was observed, and (4) Result of objective assessment supported the visual assessment. The quality of compressed images at thresholds 10, 15, and 20 was significantly better than that of input images at P < 0.0001. However, the number of unacceptable compressed images at thresholds 10, 15, and 20 was 6, 38, and 70, respectively.
CONCLUSIONS: Up to 96.65%, near-losses compression of Tc-99 m DMSA images was found using DCT by thresholding DCT coefficients at a threshold value equal to 5.
Full text links
Related Resources
Trending Papers
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Eosinophilic Esophagitis: Clinical Pearls for Primary Care Providers and Gastroenterologists.Mayo Clinic Proceedings 2024 April
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app