Active compression-decompression. A new method of cardiopulmonary resuscitation. Cardiopulmonary Resuscitation Working Group.
JAMA 1992 June 4
OBJECTIVE: To describe and compare with standard cardiopulmonary resuscitation (CPR) in humans a new form of CPR that involves both active compression and active decompression of the chest.
DESIGN: Patients in cardiac arrest in whom standard advanced cardiac life support failed were randomized to receive 2 minutes of either standard or active compression-decompression (ACD) CPR using a custom, hand-held suction device, followed by 2 minutes of the alternate technique. The ACD device was applied midsternum and used to perform CPR according to the guidelines of the American Heart Association: 80 compressions per minute, compression depth of 3.8 to 5 cm, 50% duty cycle, and constant-volume ventilation. Mechanical Thumper CPR was also compared in five patients. End-tidal carbon dioxide (ETCO2) concentrations and hemodynamic variables were measured. Transesophageal Doppler echocardiography was used to assess contractility, the velocity time integral (an analogue of cardiac output), and diastolic myocardial filling times.
RESULTS: Ten patients were enrolled. The mean +/- SD ETCO2 was 4.3 +/- 3.8 mm Hg with standard CPR and 9.0 +/- 3.9 mm Hg with ACD CPR (P less than .0001). Systolic arterial pressure with standard CPR was 52.5 +/- 14.0 mm Hg and with ACD CPR, 88.9 +/- 24.7 mm Hg (P less than .003). The velocity time integral increased from 7.3 +/- 2.6 cm with standard CPR to 17.5 +/- 5.6 cm with ACD CPR (P less than .0001), and diastolic filling times increased from 0.23 +/- .09 seconds with standard CPR to 0.37 +/- .12 seconds with ACD CPR (P less than .004). Mechanical Thumper CPR consistently underperformed both standard and ACD CPR. Minute ventilation obtained in four patients during ACD CPR without endotracheal ventilation was 6.6 +/- 0.9 L/min. After 1 hour of standard CPR failed, three of 10 patients randomized to ACD CPR rapidly converted to a hemodynamically stable rhythm following 2 minutes of ACD CPR.
CONCLUSION: ACD CPR is a simple manual technique that improved cardiopulmonary circulation in 10 patients during cardiac arrest. Although ACD CPR may have produced a return of spontaneous circulation in three patients refractory to standard measures, its impact on survival when used early in cardiac arrest remains to be determined.
DESIGN: Patients in cardiac arrest in whom standard advanced cardiac life support failed were randomized to receive 2 minutes of either standard or active compression-decompression (ACD) CPR using a custom, hand-held suction device, followed by 2 minutes of the alternate technique. The ACD device was applied midsternum and used to perform CPR according to the guidelines of the American Heart Association: 80 compressions per minute, compression depth of 3.8 to 5 cm, 50% duty cycle, and constant-volume ventilation. Mechanical Thumper CPR was also compared in five patients. End-tidal carbon dioxide (ETCO2) concentrations and hemodynamic variables were measured. Transesophageal Doppler echocardiography was used to assess contractility, the velocity time integral (an analogue of cardiac output), and diastolic myocardial filling times.
RESULTS: Ten patients were enrolled. The mean +/- SD ETCO2 was 4.3 +/- 3.8 mm Hg with standard CPR and 9.0 +/- 3.9 mm Hg with ACD CPR (P less than .0001). Systolic arterial pressure with standard CPR was 52.5 +/- 14.0 mm Hg and with ACD CPR, 88.9 +/- 24.7 mm Hg (P less than .003). The velocity time integral increased from 7.3 +/- 2.6 cm with standard CPR to 17.5 +/- 5.6 cm with ACD CPR (P less than .0001), and diastolic filling times increased from 0.23 +/- .09 seconds with standard CPR to 0.37 +/- .12 seconds with ACD CPR (P less than .004). Mechanical Thumper CPR consistently underperformed both standard and ACD CPR. Minute ventilation obtained in four patients during ACD CPR without endotracheal ventilation was 6.6 +/- 0.9 L/min. After 1 hour of standard CPR failed, three of 10 patients randomized to ACD CPR rapidly converted to a hemodynamically stable rhythm following 2 minutes of ACD CPR.
CONCLUSION: ACD CPR is a simple manual technique that improved cardiopulmonary circulation in 10 patients during cardiac arrest. Although ACD CPR may have produced a return of spontaneous circulation in three patients refractory to standard measures, its impact on survival when used early in cardiac arrest remains to be determined.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app