Ocular radiation exposure is negligible in normal volume endourological practice.
Annals of the Royal College of Surgeons of England 2024 March 7
INTRODUCTION: The annual dose limit for radiation exposure to the eye has been reduced recently; the eye is widely recognised as one of the most radiosensitive tissues in the body. There is minimal good quality research as to the radiation dose that the eye receives during endourological surgery and this study aimed to address this.
METHODS: A prospective study was performed over an 8-month period at a single large teaching hospital in the UK. Three index procedures were included: ureteric stent insertion, ureteroscopy (URS) and percutaneous nephrolithotomy (PCNL). Surgeons wore a dosimeter on the glabella with fluoroscopy time (FT) and dose area product (DAP) recorded for each case.
RESULTS: A total of 404 procedures were included (247 URSs, 150 ureteric stent insertions and 7 PCNLs). Dosimeters were worn by ten surgeons. Mean FTs (URS 20.56s; ureteric stent 18.96s; PCNL 360.67s) and mean DAP (URS 100.82cGy/m2 , ureteric stent 119.82cGy/m2 and PCNL 1121.62cGy/m2 ) were identified with significant intersurgeon variability. No surgeon had a total dosimeter dose >0.00mSv.
CONCLUSIONS: The International Commission on Radiological Protection recently reduced the yearly eye dose limit from 150 to 20mSv. Cataractogenesis is no longer considered a typical deterministic effect, with a threshold level below which no effect occurs. Even in higher volume centres, these annual limits are unlikely to be reached. Lead glasses may be considered for surgeons and radiologists with the highest exposure but, for the majority, ocular radiation exposure is negligible.
METHODS: A prospective study was performed over an 8-month period at a single large teaching hospital in the UK. Three index procedures were included: ureteric stent insertion, ureteroscopy (URS) and percutaneous nephrolithotomy (PCNL). Surgeons wore a dosimeter on the glabella with fluoroscopy time (FT) and dose area product (DAP) recorded for each case.
RESULTS: A total of 404 procedures were included (247 URSs, 150 ureteric stent insertions and 7 PCNLs). Dosimeters were worn by ten surgeons. Mean FTs (URS 20.56s; ureteric stent 18.96s; PCNL 360.67s) and mean DAP (URS 100.82cGy/m2 , ureteric stent 119.82cGy/m2 and PCNL 1121.62cGy/m2 ) were identified with significant intersurgeon variability. No surgeon had a total dosimeter dose >0.00mSv.
CONCLUSIONS: The International Commission on Radiological Protection recently reduced the yearly eye dose limit from 150 to 20mSv. Cataractogenesis is no longer considered a typical deterministic effect, with a threshold level below which no effect occurs. Even in higher volume centres, these annual limits are unlikely to be reached. Lead glasses may be considered for surgeons and radiologists with the highest exposure but, for the majority, ocular radiation exposure is negligible.
Full text links
Related Resources
Trending Papers
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Eosinophilic Esophagitis: Clinical Pearls for Primary Care Providers and Gastroenterologists.Mayo Clinic Proceedings 2024 April
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app