We have located links that may give you full text access.
First-in-human feasibility study of the aXess graft (aXess-FIH): 6-Month results.
Journal of Vascular Access 2024 Februrary 6
OBJECTIVE: The creation of an arteriovenous fistula (AVF) is considered the most effective hemodialysis (HD) vascular access. For patients who are not suitable for AVF, arteriovenous grafts (AVGs) are the best access option for chronic HD. However, conventional AVGs are prone to intimal hyperplasia, stenosis, thrombosis, and infection. Xeltis has developed an AVG as a potential alternative to currently available AVGs based on the concept of endogenous tissue restoration. The results of the first 6-month follow-up are presented here.
METHODS: The aXess first-in-human (FIH) study [NCT04898153] is a prospective, single-arm, multicenter feasibility study that evaluates the early safety and performance of the aXess Hemodialysis Graft. A total of 20 patients with end-stage renal disease were enrolled across six European investigational sites.
RESULTS: At 6-months follow-up, all grafts were patent with primary and secondary patency rates were 80% and 100%, respectively. Three patients required a re-intervention to maintain graft patency, while one re-intervention was required to restore patency. One graft thrombosis and zero infections were reported.
CONCLUSION: The expected advantages of the novel aXess Hemodialysis Graft over conventional AVGs would be evaluated by the analysis on long-term safety and effectiveness during the 5-year follow-up of the currently ongoing trial.
METHODS: The aXess first-in-human (FIH) study [NCT04898153] is a prospective, single-arm, multicenter feasibility study that evaluates the early safety and performance of the aXess Hemodialysis Graft. A total of 20 patients with end-stage renal disease were enrolled across six European investigational sites.
RESULTS: At 6-months follow-up, all grafts were patent with primary and secondary patency rates were 80% and 100%, respectively. Three patients required a re-intervention to maintain graft patency, while one re-intervention was required to restore patency. One graft thrombosis and zero infections were reported.
CONCLUSION: The expected advantages of the novel aXess Hemodialysis Graft over conventional AVGs would be evaluated by the analysis on long-term safety and effectiveness during the 5-year follow-up of the currently ongoing trial.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
Intravenous infusion of dexmedetomidine during the surgery to prevent postoperative delirium and postoperative cognitive dysfunction undergoing non-cardiac surgery: a meta-analysis of randomized controlled trials.European Journal of Medical Research 2024 April 19
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Management of Diverticulitis: A Review.JAMA Surgery 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app