We have located links that may give you full text access.
Multicenter retrospective evaluation of magnetic resonance imaging in pediatric and congenital heart disease patients with cardiac implantable electronic devices.
BACKGROUND: Guidelines addressing magnetic resonance imaging (MRI) in patients with cardiac implantable electronic devices (CIEDs) provide algorithms for imaging pediatric and congenital heart disease (CHD) patients. Guideline acceptance varies by institution. Guidelines also do not support routine MRI scans in patients with epicardial or abandoned leads, common in pediatric and CHD patients.
OBJECTIVE: The purpose of this study was to determine the incidence of MRI-related complications in pediatric and CHD patients with CIEDs, including epicardial and/or abandoned leads.
METHODS: A multicenter retrospective review included patients with CIEDs who underwent any MRI between 2007 and 2022 at congenital cardiac centers. The primary outcome was any patient adverse event or clinically significant CIED change after MRI, defined as pacing lead capture threshold increase >0.5 V with output change, P- or R- wave amplitude decrease >50% with sensitivity change, or impedance change >50%.
RESULTS: Across 14 institutions, 314 patients (median age 18.8 [1.3; 31.4] years) underwent 389 MRIs. There were 288 pacemakers (74%) and 87 implantable cardioverter-defibrillators (22%); 52% contained epicardial leads, and 14 (4%) were abandoned leads only. Symptoms or CIED changes occurred in 4.9% of MRI scans (6.1% of patients). On 9 occasions (2%), warmth or pain occurred. Pacing capture threshold or lead impedance changes occurred in 1.4% and 2.0% of CIEDs post-MRI and at follow-up.
CONCLUSION: Our data provide evidence that MRIs can be performed in pediatric and CHD patients with CIEDs, including non-MRI-conditional CIEDs and epicardial and/or abandoned leads, with rare minor symptoms or CIED changes but no other complications.
OBJECTIVE: The purpose of this study was to determine the incidence of MRI-related complications in pediatric and CHD patients with CIEDs, including epicardial and/or abandoned leads.
METHODS: A multicenter retrospective review included patients with CIEDs who underwent any MRI between 2007 and 2022 at congenital cardiac centers. The primary outcome was any patient adverse event or clinically significant CIED change after MRI, defined as pacing lead capture threshold increase >0.5 V with output change, P- or R- wave amplitude decrease >50% with sensitivity change, or impedance change >50%.
RESULTS: Across 14 institutions, 314 patients (median age 18.8 [1.3; 31.4] years) underwent 389 MRIs. There were 288 pacemakers (74%) and 87 implantable cardioverter-defibrillators (22%); 52% contained epicardial leads, and 14 (4%) were abandoned leads only. Symptoms or CIED changes occurred in 4.9% of MRI scans (6.1% of patients). On 9 occasions (2%), warmth or pain occurred. Pacing capture threshold or lead impedance changes occurred in 1.4% and 2.0% of CIEDs post-MRI and at follow-up.
CONCLUSION: Our data provide evidence that MRIs can be performed in pediatric and CHD patients with CIEDs, including non-MRI-conditional CIEDs and epicardial and/or abandoned leads, with rare minor symptoms or CIED changes but no other complications.
Full text links
Related Resources
Trending Papers
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app