Barrier Devices, Intubation, and Aerosol Mitigation Strategies: PPE in the Time of COVID-19.
Anesthesia and Analgesia 2020 September 16
BACKGROUND: Numerous barrier devices have recently been developed and rapidly deployed worldwide in an effort to protect healthcare workers (HCWs) from exposure to COVID-19 during high-risk procedures. However, only a few studies have examined their impact on the dispersion of droplets and aerosols, which are both thought to be significant contributors to the spread of COVID-19.
METHODS: Two commonly used barrier devices, an intubation box and a clear plastic intubation sheet, were evaluated using a physiologically accurate cough simulator. Aerosols were modeled using a commercially available fog machine and droplets were modeled with fluorescein dye. Both particles were propelled by the cough simulator in a simulated intubation environment. Data was captured by high-speed flash photography and aerosol and droplet dispersion were assessed qualitatively with and without a barrier in place.
RESULTS: Droplet contamination after a simulated cough was seemingly contained by both barrier devices. Simulated aerosol escaped the barriers and flowed towards the head of the bed. During barrier removal, simulated aerosol trapped underneath was released and propelled towards the HCW at the head of the bed. Usage of the intubation sheet concentrated droplets onto a smaller area. If no barrier was used, positioning the patient in slight reverse Trendelenburg directed aerosols away from the healthcare worker located at the head of the bed.
CONCLUSION: Our observations imply that intubation boxes and sheets may reduce HCW exposure to droplets, but they both may merely redirect aerosolized particles, potentially resulting in increased exposure to aerosols in certain circumstances. Aerosols may remain within the barrier device after a cough and manipulation of the box may release them. Patients should be positioned to facilitate intubation, but slight reverse Trendelenburg may direct infectious aerosols away from the HCW. Novel barrier devices should be used with caution and further validation studies are necessary.
METHODS: Two commonly used barrier devices, an intubation box and a clear plastic intubation sheet, were evaluated using a physiologically accurate cough simulator. Aerosols were modeled using a commercially available fog machine and droplets were modeled with fluorescein dye. Both particles were propelled by the cough simulator in a simulated intubation environment. Data was captured by high-speed flash photography and aerosol and droplet dispersion were assessed qualitatively with and without a barrier in place.
RESULTS: Droplet contamination after a simulated cough was seemingly contained by both barrier devices. Simulated aerosol escaped the barriers and flowed towards the head of the bed. During barrier removal, simulated aerosol trapped underneath was released and propelled towards the HCW at the head of the bed. Usage of the intubation sheet concentrated droplets onto a smaller area. If no barrier was used, positioning the patient in slight reverse Trendelenburg directed aerosols away from the healthcare worker located at the head of the bed.
CONCLUSION: Our observations imply that intubation boxes and sheets may reduce HCW exposure to droplets, but they both may merely redirect aerosolized particles, potentially resulting in increased exposure to aerosols in certain circumstances. Aerosols may remain within the barrier device after a cough and manipulation of the box may release them. Patients should be positioned to facilitate intubation, but slight reverse Trendelenburg may direct infectious aerosols away from the HCW. Novel barrier devices should be used with caution and further validation studies are necessary.
Full text links
Related Resources
Trending Papers
British Society of Gastroenterology guidelines for the management of hepatocellular carcinoma in adults.Gut 2024 April 17
Systemic lupus erythematosus.Lancet 2024 April 18
Should renin-angiotensin system inhibitors be held prior to major surgery?British Journal of Anaesthesia 2024 May
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Colorectal polypectomy and endoscopic mucosal resection: European Society of Gastrointestinal Endoscopy (ESGE) Guideline - Update 2024.Endoscopy 2024 April 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app