We have located links that may give you full text access.
Effects of MRI scanner parameters on breast cancer radiomics.
Expert Systems with Applications 2017 November 31
Purpose: To assess the impact of varying magnetic resonance imaging (MRI) scanner parameters on the extraction of algorithmic features in breast MRI radiomics studies.
Methods: In this retrospective study, breast imaging data for 272 patients were analyzed with magnetic resonance (MR) images. From the MR images, we assembled and implemented 529 algorithmic features of breast tumors and fibrograndular tissue (FGT). We divided the features into 10 groups based on the type of data used for the feature extraction and the nature of the extracted information. Three scanner parameters were considered: scanner manufacturer, scanner magnetic field strength, and slice thickness. We assessed the impact of each of the scanner parameters on each of the feature by testing whether the feature values are systematically diverse for different values of these scanner parameters. A two-sample t-test has been used to establish whether the impact of a scanner parameter on values of a feature is significant and receiver operating characteristics have been used for to establish the extent of that effect.
Results: On average, higher proportion (69% FGT versus 20% tumor) of FGT related features were affected by the three scanner parameters. Of all feature groups and scanner parameters, the feature group related to the variation in FGT enhancement was found to be the most sensitive to the scanner manufacturer (AUC = 0.81 ± 0.14).
Conclusions: Features involving calculations from FGT are particularly sensitive to the scanner parameters.
Methods: In this retrospective study, breast imaging data for 272 patients were analyzed with magnetic resonance (MR) images. From the MR images, we assembled and implemented 529 algorithmic features of breast tumors and fibrograndular tissue (FGT). We divided the features into 10 groups based on the type of data used for the feature extraction and the nature of the extracted information. Three scanner parameters were considered: scanner manufacturer, scanner magnetic field strength, and slice thickness. We assessed the impact of each of the scanner parameters on each of the feature by testing whether the feature values are systematically diverse for different values of these scanner parameters. A two-sample t-test has been used to establish whether the impact of a scanner parameter on values of a feature is significant and receiver operating characteristics have been used for to establish the extent of that effect.
Results: On average, higher proportion (69% FGT versus 20% tumor) of FGT related features were affected by the three scanner parameters. Of all feature groups and scanner parameters, the feature group related to the variation in FGT enhancement was found to be the most sensitive to the scanner manufacturer (AUC = 0.81 ± 0.14).
Conclusions: Features involving calculations from FGT are particularly sensitive to the scanner parameters.
Full text links
Related Resources
Trending Papers
Guillain-Barré syndrome: History, pathogenesis, treatment, and future directions.European Journal of Neurology 2024 May 17
Contrast-induced acute kidney injury: a review of definition, pathogenesis, risk factors, prevention and treatment.BMC Nephrology 2024 April 23
Angiotensin Receptor Blocker-Neprilysin Inhibitor for Heart Failure with Reduced Ejection Fraction.Pharmacological Research : the Official Journal of the Italian Pharmacological Society 2024 May 12
The Therapy and Management of Heart Failure with Preserved Ejection Fraction: New Insights on Treatment.Cardiac Failure Review 2024
European Respiratory Society Clinical Practice Guideline on symptom management for adults with serious respiratory illness.European Respiratory Journal 2024 May 9
Axillary Surgery for Breast Cancer in 2024.Cancers 2024 April 24
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app