An inflatable balloon catheter and liquid 125I radiation source (GliaSite Radiation Therapy System) for treatment of recurrent malignant glioma: multicenter safety and feasibility trial.

OBJECT: In this study the authors evaluated the safety and performance of the GliaSite Radiation Therapy System (RTS) in patients with recurrent malignant brain tumors who were undergoing tumor resection.

METHODS: The GliaSite is an inflatable balloon catheter that is placed in the resection cavity at the time of tumor debulking. Low-dose-rate radiation is delivered with an aqueous solution of organically bound iodine-125 (lotrex [sodium 3-(125I)-iodo-4-hydroxybenzenesulfonate]), which are temporarily introduced into the balloon portion of the device via a subcutaneous port. Adults with recurrent malignant glioma underwent resection and GliaSite implantation. One to 2 weeks later, the device was filled with Iotrex for 3 to 6 days, following which the device was explanted. Twenty-one patients with recurrent high-grade astrocytomas were enrolled in the study and received radiation therapy. There were two end points: 1) successful implantation and delivery of brachytherapy; and 2) safety of the device. Implantation of the device, delivery of radiation, and the explantation procedure were well tolerated. At least 40 to 60 Gy was delivered to all tissues within the target volume. There were no serious adverse device-related events during brachytherapy. One patient had a pseudomeningocele, one patient had a wound infection, and three patients had meningitis (one bacterial, one chemical, and one aseptic). No symptomatic radiation necrosis was identified during 21.8 patient-years of follow up. The median survival of previously treated patients was 12.7 months (95% confidence interval 6.9-15.3 months).

CONCLUSIONS: The GliaSite RTS performs safely and efficiently. It delivers a readily quantifiable dose of radiation to tissue at the highest risk for tumor recurrence.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app