We have located links that may give you full text access.
Invasive monitoring of limbic epilepsy using stereotactic depth and subdural strip electrodes: surgical technique.
Surgical Neurology 1997 July
BACKGROUND: In spite of advances in noninvasive localization of seizure foci, some cases of intractable limbic epilepsy still require invasive recordings in order to identify the site of seizure onset. This necessitates a safe and reliable method for placing depth and subdural electrodes in the mesial temporal and orbitofrontal regions. The University of Florida has devised a system that utilizes CRW-based stereotactic placement of bitemporal depth electrodes in conjunction with placement of subdural strips over the inferolateral temporal lobe and orbitofrontal cortex. This report describes the surgical technique and initial clinical experience using this method.
METHODS: Depth electrodes are placed along the long axis of the hippocampi via an occipital approach. A CRW-based stereotactic system was developed that incorporates both computed tomography (CT) and magnetic resonance imaging (MRI) for selection of target and entry sites and displaying the electrode trajectory. Subdural strip electrodes are placed over the inferolateral temporal and orbitofrontal regions through burr holes.
RESULTS: This method has been used in 18 patients (depth electrodes only in three patients and depth electrodes with subdural strips in 15 patients). This information lead to surgical resections in 15 patients. No resection was recommended in three patients (two with bitemporal onset and one with no seizures after 6 weeks). Complications were limited to an unplanned removal of one electrode and an asymptomatic lateral temporal lobe contusion in one patient.
CONCLUSIONS: This method provides a safe and effective way to sample bilateral mesial temporal and orbitofrontal regions in cases of intractable limbic epilepsy.
METHODS: Depth electrodes are placed along the long axis of the hippocampi via an occipital approach. A CRW-based stereotactic system was developed that incorporates both computed tomography (CT) and magnetic resonance imaging (MRI) for selection of target and entry sites and displaying the electrode trajectory. Subdural strip electrodes are placed over the inferolateral temporal and orbitofrontal regions through burr holes.
RESULTS: This method has been used in 18 patients (depth electrodes only in three patients and depth electrodes with subdural strips in 15 patients). This information lead to surgical resections in 15 patients. No resection was recommended in three patients (two with bitemporal onset and one with no seizures after 6 weeks). Complications were limited to an unplanned removal of one electrode and an asymptomatic lateral temporal lobe contusion in one patient.
CONCLUSIONS: This method provides a safe and effective way to sample bilateral mesial temporal and orbitofrontal regions in cases of intractable limbic epilepsy.
Full text links
Related Resources
Trending Papers
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app