Beware of atrial pacing-induced ventricular depolarization: A case of lead malfunction.
Pacing and Clinical Electrophysiology : PACE 2024 April 30
Although ventricular capture during the atrial threshold test is possible, there are rare reports on the insulation defect and inactive leads thereof. In this case, we present a pacemaker-dependent patient with a history of pacemaker generator replacements. The patient experienced ventricular capture induced by atrial pacing due to adhesion of the atrial and ventricular leads with an insulation defect. The atrial lead was abandoned and a new lead was implanted. However, there was a significant decrease in ventricular impedance detected shortly after the new lead was implanted. When observing the phenomenon of atrial pacing-induced ventricular depolarization, one uncommon reason to consider is lead adhesive wear. It is important to pay attention to the contact and bending sites of the leads.
Full text links
Related Resources
Trending Papers
Obesity pharmacotherapy in older adults: a narrative review of evidence.International Journal of Obesity 2024 May 7
SGLT2 Inhibitors in Kidney Diseases-A Narrative Review.International Journal of Molecular Sciences 2024 May 2
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app