We have located links that may give you full text access.
Utilizing collimated aperture with proton pencil beam scanning (PBS) for stereotactic radiotherapy.
Journal of Applied Clinical Medical Physics 2024 April 27
PURPOSE: Proton stereotactic radiosurgery (PSRS) has emerged as an innovative proton therapy modality aimed at achieving precise dose delivery with minimal impact on healthy tissues. This study explores the dosimetric outcomes of PSRS in comparison to traditional intensity-modulated proton therapy (IMPT) by focusing on cases with small target volumes. A custom-made aperture system designed for proton therapy, specifically tailored to small target volumes, was developed and implemented for this investigation.
METHODS: A prerequisite mechanical validation through an isocentricity test precedes dosimetric assessments, ensuring the seamless integration of mechanical and dosimetry analyses. Five patients were enrolled in the study, including two with choroid melanoma and three with arteriovenous malformations (AVM). Two treatment plans were meticulously executed for each patient, one utilizing a collimated aperture and the other without. Both plans were subjected to robust optimization, maintaining identical beam arrangements and consistent optimization parameters to account for setup errors of 2 mm and range uncertainties of 3.5%. Plan evaluation metrics encompassing the Heterogeneity Index (HI), Paddick Conformity Index (CIPaddick ), Gradient Index (GI), and the R50% index to evaluate alterations in low-dose volume distribution.
RESULTS: The comparative analysis between PSRS and traditional PBS treatment revealed no significant differences in plan outcomes, with both modalities demonstrating comparable target coverage. However, collimated apertures resulted in discernible improvements in dose conformity, dose fall-off, and reduced low-dose volume.
CONCLUSIONS: This study underscores the advantageous impact of the aperture system on proton therapy, particularly in cases involving small target volumes.
METHODS: A prerequisite mechanical validation through an isocentricity test precedes dosimetric assessments, ensuring the seamless integration of mechanical and dosimetry analyses. Five patients were enrolled in the study, including two with choroid melanoma and three with arteriovenous malformations (AVM). Two treatment plans were meticulously executed for each patient, one utilizing a collimated aperture and the other without. Both plans were subjected to robust optimization, maintaining identical beam arrangements and consistent optimization parameters to account for setup errors of 2 mm and range uncertainties of 3.5%. Plan evaluation metrics encompassing the Heterogeneity Index (HI), Paddick Conformity Index (CIPaddick ), Gradient Index (GI), and the R50% index to evaluate alterations in low-dose volume distribution.
RESULTS: The comparative analysis between PSRS and traditional PBS treatment revealed no significant differences in plan outcomes, with both modalities demonstrating comparable target coverage. However, collimated apertures resulted in discernible improvements in dose conformity, dose fall-off, and reduced low-dose volume.
CONCLUSIONS: This study underscores the advantageous impact of the aperture system on proton therapy, particularly in cases involving small target volumes.
Full text links
Related Resources
Trending Papers
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app