Development and validation of ivermectin quantification method in volumetric absorptive microsampling using liquid chromatography-tandem mass spectrometry.

Heliyon 2024 April 31
BACKGROUND: Ivermectin is a broad-spectrum anthelmintic used to control onchocerciasis from nematode parasites. As an anthelmintic, ivermectin is designed to have high levels in the gastrointestinal tract, so that the systemic intake is relatively low. Due to the very small concentration of ivermectin, a selective and sensitive approach is needed for the analysis of ivermectin in blood. Several methods have been developed using plasma and Dried Blood Spots, but there are still shortcomings due to hematocrit effects. Therefore, this study was conducted to establish a validated ivermectin analysis method with doramectin as the internal standard in using Ultra High-Performance Liquid Chromatography-Tandem Mass Spectrometry.

METHODS: Mass spectrometry equipped with triple quadrupole and positive electrospray ionization mode was used to conduct the analysis. For the biological matrix, whole blood was used by Volumetric Absorptive Microsampling and extracted using a protein precipitation technique with a combination of acetonitrile and methanol (1:1). VAMS has some advantages such as not being affected by hematocrit, requires a small and fixed volume of sample, also a more efficient sampling process.

RESULTS: The optimum conditions were achieved with an Acquity® UPLC BEH C18 column (1,7 μm; 2.1 × 100 mm); extracted-flow rate was 0,2 mL/min; mobile phase was 5 mM ammonium formate pH 3.00 and acetonitrile (10:90) with isocratic elution. Multiple Reaction Monitoring (MRM) detection by m / z values was 892.41 > 569.5 for ivermectin and 916,41 > 331,35 for doramectin.

CONCLUSION: The method has been appropriately validated in compliance with the 2018 guidelines laid out by the US Food and Drug Administration. Resulting the minimum detection (LLOQ) was 1 ng/mL with a linear concentration range spanning from 1 to 150 ng/mL.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app