Add like
Add dislike
Add to saved papers

Radiation Exposure and Contrast Agent Use during Endovascular Aortic Repair Using Mobile Versus Fixed Angiography Systems.

BACKGROUND: For (thoracic) endovascular aortic repair ((T)EVAR) procedures, both mobile (standard operating room (SOR)) and fixed C-arm (hybrid operating room (HOR)) systems are available. This study evaluated differences in key procedural parameters, and procedural success for (T)EVAR in the SOR versus the HOR.

METHODS: All patients who underwent standard elective (T)EVAR at the Clinic for Vascular and Endovascular Surgery at the University Hospital Duesseldorf, Germany, between 1 January 2012 and 1 January 2019 were included. Data were retrieved from archived medical records. Endpoints were analyzed for SOR versus HOR during (T)EVAR.

RESULTS: A total of 93 patients, including 50 EVAR (SOR ( n = 20); HOR ( n = 30)) and 43 TEVAR (SOR ( n = 22); HOR ( n = 21)) were included. The dose area product (DAP) for EVAR and TEVAR was lower in the SOR than in the HOR (EVAR, SOR: 1635 ± 1088 cGy·cm2 ; EVAR, HOR: 7819 ± 8928 cGy·cm2 ; TEVAR, SOR: 8963 ± 34,458 cGy·cm2 ; TEVAR, HOR: 14,591 ± 11,584 cGy·cm2 ( p < 0.05)). Procedural fluoroscopy time was shorter in the SOR than in the HOR for EVAR and TEVAR (EVAR, SOR: 7 ± 4 min; EVAR, HOR: 18.8 ± 11.3 min; TEVAR, SOR: 6.6 ± 9.6 min; TEVAR, HOR: 13.9 ± 11.8 min ( p < 0.05)). Higher volumes of contrast agent were applied during EVAR and TEVAR in the SOR than in the HOR (EVAR, SOR: 57.5 ± 20 mL; EVAR: HOR: 33.3 ± 5 mL ( p < 0.05); TEVAR; SOR: 71.5 ± 53.4 mL, TEVAR, HOR: 48.2 ± 27.5 mL ( p ≥ 0.05).

CONCLUSION: The use of a fixed C-arm angiography system in the HOR results in higher radiation exposure and longer fluoroscopy times but lower contrast agent volumes when compared with mobile C-arm systems in the SOR. Because stochastic radiation sequelae are more likely to be tolerated in an older patient population and, in addition, there is a higher incidence of CKD in this patient population, allocation of patients to the HOR for standard (T)EVAR seems particularly advisable based on our results.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app