BACKGROUND AND OBJECTIVES: Robot-assisted biopsies have gained popularity in the last years. Most robotic procedures are performed with a floor-based robotic arm. Recently, Medtronic Stealth Autoguide, a miniaturized robotic arm that work together with an optical neuronavigation system, was launched. Its application in pediatric cases is relatively unexplored. In this study, we retrospectively report our experience using the Stealth Autoguide, for frameless stereotactic biopsies in pediatric patients.
METHODS: Pediatric patients who underwent stereotactic biopsy using the Stealth Autoguide cranial robotic platform from July 2020 to May 2023 were included in this study. Clinical, neuroradiological, surgical, and histological data were collected and analyzed.
RESULTS: Nineteen patients underwent 20 procedures (mean age was 9-year-old, range 1-17). In four patients, biopsy was part of a more complex surgical procedure (laser interstitial thermal therapy - LITT). The most common indication was diffuse intrinsic brain stem tumor, followed by diffuse supratentorial tumor. Nine procedures were performed in prone position, eight in supine position, and three in lateral position. Facial surface registration was adopted in six procedures, skull-fixed fiducials in 14. The biopsy diagnostic tissue acquisition rate was 100% in the patients who underwent only biopsy, while in the biopsy/LITT group, one case was not diagnostic. No patients developed clinically relevant postoperative complications.
CONCLUSION: The Stealth Autoguide system has proven to be safe, diagnostic, and highly accurate in performing stereotactic biopsies for both supratentorial and infratentorial lesions in the pediatric population.
METHODS: Pediatric patients who underwent stereotactic biopsy using the Stealth Autoguide cranial robotic platform from July 2020 to May 2023 were included in this study. Clinical, neuroradiological, surgical, and histological data were collected and analyzed.
RESULTS: Nineteen patients underwent 20 procedures (mean age was 9-year-old, range 1-17). In four patients, biopsy was part of a more complex surgical procedure (laser interstitial thermal therapy - LITT). The most common indication was diffuse intrinsic brain stem tumor, followed by diffuse supratentorial tumor. Nine procedures were performed in prone position, eight in supine position, and three in lateral position. Facial surface registration was adopted in six procedures, skull-fixed fiducials in 14. The biopsy diagnostic tissue acquisition rate was 100% in the patients who underwent only biopsy, while in the biopsy/LITT group, one case was not diagnostic. No patients developed clinically relevant postoperative complications.
CONCLUSION: The Stealth Autoguide system has proven to be safe, diagnostic, and highly accurate in performing stereotactic biopsies for both supratentorial and infratentorial lesions in the pediatric population.
Full text links
Related Resources
Trending Papers
Haemodynamic monitoring during noncardiac surgery: past, present, and future.Journal of Clinical Monitoring and Computing 2024 April 31
2024 AHA/ACC/AMSSM/HRS/PACES/SCMR Guideline for the Management of Hypertrophic Cardiomyopathy: A Report of the American Heart Association/American College of Cardiology Joint Committee on Clinical Practice Guidelines.Circulation 2024 May 9
Obesity pharmacotherapy in older adults: a narrative review of evidence.International Journal of Obesity 2024 May 7
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app