Mechanics of dynamic compression plate application in fracture fixation.
Clinical Biomechanics 2024 Februrary 17
BACKGROUND: Dynamic compression plating is a fundamental type of bone fracture fixation used to generate interfragmentary compression. The goal of this study was to investigate the mechanics of the surgical application of these plates, specifically how plate prebend, screw location, fracture gap, and applied torque influence the resulting compressive pressures.
METHODS: Synthetic bones with transverse fractures were fixed with locking compression plates. One side of the fracture was fixed with locking screws. On the other side of the fracture, a nonlocking screw was inserted eccentrically to induce interfragmentary compression. A pressure mapping sensor within the fracture gap was used to record the resulting pressure distribution. Plate prebends of 0 mm, 1.5 mm, and 3 mm were tested. Three locations of the eccentric screw, four levels of screw torque, and two initial fracture gap conditions also were tested.
FINDINGS: With increasing plate prebend, fracture compression pressures shifted significantly toward the far cortex; however, compression force decreased (P < 0.05). The 1.5 mm prebend plate resulted in the greatest contact area. Increasing screw torque generally resulted in greater fracture compression force. The introduction of a 1 mm fracture gap at the far cortex prior to dynamic compression resulted in little or no fracture compression.
INTERPRETATION: The model showed that increasing plate prebend results in an increasing shift of fracture compression pressures toward the far cortex; however, this is accompanied by decreases in compressive force. Initial fracture gaps at the far cortex can result in little or no compression.
METHODS: Synthetic bones with transverse fractures were fixed with locking compression plates. One side of the fracture was fixed with locking screws. On the other side of the fracture, a nonlocking screw was inserted eccentrically to induce interfragmentary compression. A pressure mapping sensor within the fracture gap was used to record the resulting pressure distribution. Plate prebends of 0 mm, 1.5 mm, and 3 mm were tested. Three locations of the eccentric screw, four levels of screw torque, and two initial fracture gap conditions also were tested.
FINDINGS: With increasing plate prebend, fracture compression pressures shifted significantly toward the far cortex; however, compression force decreased (P < 0.05). The 1.5 mm prebend plate resulted in the greatest contact area. Increasing screw torque generally resulted in greater fracture compression force. The introduction of a 1 mm fracture gap at the far cortex prior to dynamic compression resulted in little or no fracture compression.
INTERPRETATION: The model showed that increasing plate prebend results in an increasing shift of fracture compression pressures toward the far cortex; however, this is accompanied by decreases in compressive force. Initial fracture gaps at the far cortex can result in little or no compression.
Full text links
Related Resources
Trending Papers
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Finerenone: From the Mechanism of Action to Clinical Use in Kidney Disease.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app