Automatic adjustment of oxygen concentration during high-flow nasal cannula treatment using a targeted SpO2 feedback system.

Nursing in Critical Care 2024 Februrary 7
AIM: To test whether targeted SpO2 feedback (TSF), an automatic control system for fraction of inspired oxygen (FiO2), achieves more time in the optimal SpO2 range and/or reduces the frequency of manual adjustments to administered FiO2 compared with conventional manual titration in patients with hypoxia on high-flow nasal cannula (HFNC) therapy.

STUDY DESIGN: Twenty-two patients were recruited from two hospitals. For each, two sessions of manual mode and two sessions of TSF were applied in a random order, each session lasting 2 h. The target SpO2 on TSF was 95%. Oxygen monitoring levels were classified into four SpO2 ranges: hypoxia (≤ 89%), borderline (90%-93%), optimal (94%-96%) and hyperoxia (≥ 97%). The two modes were compared based on the proportion of time spent in each SpO2 range and the number of manual FiO2 adjustments.

RESULTS: The proportion of time in the optimal SpO2 range was 20.5% under manual titration mode and 65.4% under TSF (p < .01). The proportions of time in the hypoxia range were 1.1% and 0.4%, respectively (p = .31), in the borderline range 4.7% and 3.5%, respectively (p = .54), and in the hyperoxia range 73.7% and 30.7%, respectively (p < .01). There were statistical differences only in the optimal and hyperoxia SpO2 ranges. During the 8 h, the frequency of manual FiO2 adjustment was 0.7 times for the manual mode and 0.2 times for TSF, showing no statistically significant difference (p = 0.076).

CONCLUSION: Compared with manual titration, TSF achieved greater time of the optimal SpO2 and less time of hyperoxia during HFNC. The frequency of manual adjustments on TSF tended to be less than on manual titration mode.

RELEVANCE TO CLINICAL PRACTICE: Automatic closed-loop algorithm FiO2 monitoring systems can achieve better oxygen treatments than conventional monitoring and may reduce nurse workloads. In the era of pandemic respiratory diseases, this system can also facilitate contactless SpO2 monitoring during HFNC therapy.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app