Impact of lead position on tricuspid regurgitation, ventricular function, and heart failure exacerbation and mortality after cardiac implantable electronic device implantation. Preliminary results from the PACE-RVTR Registry.

BACKGROUND: The most frequent mechanism of lead-related tricuspid regurgitation (LRTR), which occurs in 7.2% to 44.7% of patients implanted with a cardiac implantable electronic device (CIED), is leaflet impingement or the restriction of its movement by a ventricular lead. It is unclear if the position of the lead tip - in the right ventricular apex (RVA) or other location (non-RVA) - has any influence on the development of LRTR. The study aimed to determine the impact of the CIED lead tip position on the development or progression of tricuspid regurgitation (TR) and its potential impact on heart failure exacerbation and mortality.

METHODS: One hundred and two consecutive patients who received CIEDs between March 2020 and October 2021 were included in the prospective registry (PACE-RVTR). Patients were assigned to two groups depending on the lead position - the RVA group and the non-RVA group. All patients underwent echocardiographic evaluation before implantation and one year later.

RESULTS: In terms of baseline clinical characteristics, the two groups did not differ. Before CIED implantation, patients in the non-RVA group had better left ventricular systolic function (P = 0.004). Pacemakers were implanted more often in the non-RVA group (P = 0.001) while implantable cardioverter-defibrillators in the RVA group (P = 0.008). Progression to severe or massive TR was more common in the non-RVA group (P = 0.005).

CONCLUSION: Severe and massive TR occurred more often in patients with the non-RVA position of the lead. The right ventricular lead position did not impact heart failure progression or all-cause mortality at two-year follow-up.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app