Aortic Root Thrombosis in patients with HeartMate 3 left ventricular assist device support.
Journal of Heart and Lung Transplantation 2023 September 21
BACKGROUND: Aortic root thrombosis(ART) is a complication of continuous-flow left ventricular assist device therapy. However, the incidence and related complications of ART in HeartMate 3 (HM3) patients remain unknown.
METHODS: Patients who underwent HM3 implantation from November 2014 to August 2020 at a quaternary academic medical center were included. Demographics and outcomes were abstracted from the medical record. Echocardiograms and contrast-enhanced computed tomography studies were reviewed to identify patients who developed ART and/or moderate or greater aortic insufficiency (AI) on HM3 support.
RESULTS: The study cohort included 197 HM3 patients with a median postimplant follow-up of 17.5 months. Nineteen patients (9.6%) developed ART during HM3 support, and 15 patients (7.6%) developed moderate or greater AI. Baseline age, gender, race, implantation strategy, and INTERMACS classification were similar between the ART and no-ART groups. ART was associated with an increased risk of death, stroke, or aortic valve (AV) intervention (subhazard ratio [SHR] 3.60 [95% confidence interval (CI) 1.71-7.56]; p = 0.001) and moderate or greater AI (SHR 11.1 [CI 3.60-34.1]; p < 0.001) but was not associated with a statistically significantly increased risk of death or stroke on HM3 support (2.12 [0.86-5.22]; p = 0.10). Of the 19 patients with ART, 6 (31.6%) developed moderate or greater AI, necessitating more frequent AV interventions (ART: 5 AV interventions [3 surgical repairs, 1 surgical replacement, 1 transcatheter replacement; 26.3%]; no-ART: 0).
CONCLUSIONS: Nearly 10% of HM3 patients developed ART during device support. ART was associated with increased risk of a composite end-point of death, stroke, or AV intervention as well as moderate or greater AI.
METHODS: Patients who underwent HM3 implantation from November 2014 to August 2020 at a quaternary academic medical center were included. Demographics and outcomes were abstracted from the medical record. Echocardiograms and contrast-enhanced computed tomography studies were reviewed to identify patients who developed ART and/or moderate or greater aortic insufficiency (AI) on HM3 support.
RESULTS: The study cohort included 197 HM3 patients with a median postimplant follow-up of 17.5 months. Nineteen patients (9.6%) developed ART during HM3 support, and 15 patients (7.6%) developed moderate or greater AI. Baseline age, gender, race, implantation strategy, and INTERMACS classification were similar between the ART and no-ART groups. ART was associated with an increased risk of death, stroke, or aortic valve (AV) intervention (subhazard ratio [SHR] 3.60 [95% confidence interval (CI) 1.71-7.56]; p = 0.001) and moderate or greater AI (SHR 11.1 [CI 3.60-34.1]; p < 0.001) but was not associated with a statistically significantly increased risk of death or stroke on HM3 support (2.12 [0.86-5.22]; p = 0.10). Of the 19 patients with ART, 6 (31.6%) developed moderate or greater AI, necessitating more frequent AV interventions (ART: 5 AV interventions [3 surgical repairs, 1 surgical replacement, 1 transcatheter replacement; 26.3%]; no-ART: 0).
CONCLUSIONS: Nearly 10% of HM3 patients developed ART during device support. ART was associated with increased risk of a composite end-point of death, stroke, or AV intervention as well as moderate or greater AI.
Full text links
Related Resources
Trending Papers
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Finerenone: From the Mechanism of Action to Clinical Use in Kidney Disease.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app