How Electrode Position Affects Selective His Bundle Capture: A Modelling Study.
IEEE Transactions on Bio-medical Engineering 2021 April 10
In certain cardiac conduction system pathologies, like bundle branch block, block may be proximal, allowing for electrical stimulation of the more disal His bundle to most effectively restore activation. While selective stimulation of the His bundle is sought, surrounding myocardium may also be excited, resulting in nonselective pacing. The myocardium and His bundle have distinct capture thresholds, but the factors affecting whether His bundle pacing is selective or nonselective remain unelucidated.
OBJECTIVE: We investigated the properties which affect the capture thresholds in order to improve selective pacing.
METHODS: We performed biophysically detailed, computer simulations of a His fibre running through a septal wedge preparation to compute capture thresholds under various configurations of electrode polarity and orientation.
RESULTS: The myocardial capture threshold is close to that of the His bundle. The His fibre needed to intersect with the electrode tip to favor its activation. Inserting the electrode fully within the septum increased the myocardial capture threshold. Reversing polarity, to rely on anode break excitation, also increased the ease of selective pacing.
CONCLUSION: Model results were consistent with clinical observations. For selective pacing, the tip needs to be in contact with the His fibre and anodal stimulation is preferable.
SIGNIFICANCE: This study provides insight into helping establish electrode and stimulation parameters for selective His bundle pacing in patients.
OBJECTIVE: We investigated the properties which affect the capture thresholds in order to improve selective pacing.
METHODS: We performed biophysically detailed, computer simulations of a His fibre running through a septal wedge preparation to compute capture thresholds under various configurations of electrode polarity and orientation.
RESULTS: The myocardial capture threshold is close to that of the His bundle. The His fibre needed to intersect with the electrode tip to favor its activation. Inserting the electrode fully within the septum increased the myocardial capture threshold. Reversing polarity, to rely on anode break excitation, also increased the ease of selective pacing.
CONCLUSION: Model results were consistent with clinical observations. For selective pacing, the tip needs to be in contact with the His fibre and anodal stimulation is preferable.
SIGNIFICANCE: This study provides insight into helping establish electrode and stimulation parameters for selective His bundle pacing in patients.
Full text links
Related Resources
Trending Papers
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Revascularization Strategy in Myocardial Infarction with Multivessel Disease.Journal of Clinical Medicine 2024 March 27
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app