Verification of the Heska Element Point-of-Care blood gas instrument for use with venous blood from alpacas and llamas, with determination of reference intervals.
Veterinary Clinical Pathology 2018 September
BACKGROUND: The Heska Element POC ("EPOC") is a blood gas instrument intended for use with canine, feline, and equine whole blood; no verification for use with camelid specimens has been reported.
OBJECTIVES: Using camelid specimens and commercial quality control materials (QCM), we investigatee EPOC analytical performance and establish EPOC camelid reference intervals (RIs).
METHODS: Camelid blood (n = 124) was analyzed using the EPOC (pH, pCO2 , pO2 , HCO3 , base excess, SO2 , sodium, potassium, chloride, ionized calcium, TCO2 , anion gap, HCT, HGB, glucose, lactate, and creatinine); plasma was analyzed using a Roche Cobas c501 (sodium, potassium, chloride, TCO2 , anion gap, glucose, and creatinine). Method comparison data were assessed using Pearson's correlation, Passing-Bablok regression, and Bland-Altman plots. EPOC precision was evaluated using QCM and camelid blood.
RESULTS: For all measurands except anion gap, the EPOC vs Cobas instrument correlation was r > .85. Except for pO2 and pCO2 , EPOC precision (QCM and blood) ranged from a repeatability CV <1%-6.3%. Mild constant bias for chloride, glucose, TCO2 , anion gap, and creatinine, and mild proportional bias for chloride, glucose, and anion gap were present. The total error (QCM data) for the EPOC instrument was below the ASVCP-recommended allowable total error. Alpacas had higher potassium and lactate, while llamas had higher creatinine, sodium, chloride, ionized calcium, pO2 , and SO2 . Statistical RIs based on alpaca (n = 74-96) and llama data (n = 12-17) are reported descriptively.
CONCLUSIONS: The EPOC analyzer shows good performance with camelid blood. A lack of complete agreement with automated chemistry analyzers highlights the importance of interpreting patient data using instrument-specific RIs.
OBJECTIVES: Using camelid specimens and commercial quality control materials (QCM), we investigatee EPOC analytical performance and establish EPOC camelid reference intervals (RIs).
METHODS: Camelid blood (n = 124) was analyzed using the EPOC (pH, pCO2 , pO2 , HCO3 , base excess, SO2 , sodium, potassium, chloride, ionized calcium, TCO2 , anion gap, HCT, HGB, glucose, lactate, and creatinine); plasma was analyzed using a Roche Cobas c501 (sodium, potassium, chloride, TCO2 , anion gap, glucose, and creatinine). Method comparison data were assessed using Pearson's correlation, Passing-Bablok regression, and Bland-Altman plots. EPOC precision was evaluated using QCM and camelid blood.
RESULTS: For all measurands except anion gap, the EPOC vs Cobas instrument correlation was r > .85. Except for pO2 and pCO2 , EPOC precision (QCM and blood) ranged from a repeatability CV <1%-6.3%. Mild constant bias for chloride, glucose, TCO2 , anion gap, and creatinine, and mild proportional bias for chloride, glucose, and anion gap were present. The total error (QCM data) for the EPOC instrument was below the ASVCP-recommended allowable total error. Alpacas had higher potassium and lactate, while llamas had higher creatinine, sodium, chloride, ionized calcium, pO2 , and SO2 . Statistical RIs based on alpaca (n = 74-96) and llama data (n = 12-17) are reported descriptively.
CONCLUSIONS: The EPOC analyzer shows good performance with camelid blood. A lack of complete agreement with automated chemistry analyzers highlights the importance of interpreting patient data using instrument-specific RIs.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
Intravenous infusion of dexmedetomidine during the surgery to prevent postoperative delirium and postoperative cognitive dysfunction undergoing non-cardiac surgery: a meta-analysis of randomized controlled trials.European Journal of Medical Research 2024 April 19
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Systemic lupus erythematosus.Lancet 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app