Development of a Mock Extracorporeal Membrane Oxygenation Circuit to Assess Recirculation.
ASAIO Journal : a Peer-reviewed Journal of the American Society for Artificial Internal Organs 2016 July
A limitation of veno-venous extracorporeal membrane oxygenation (VV-ECMO) is the recirculating of newly oxygenated blood back to the VV-ECMO circuit. We compared cannulation strategies for combating recirculation utilizing a mock circulation loop (MCL) with ultrasonic flow probes and pressure transducers ensuring that the MCL was run at physiological hemodynamic parameters. Mean recirculation percentages were lower (ANOVA, F = 14.25; p = 0.0001) with the Wang-Zwische dual lumen cannula (4.00 ± 1.77, n = 7) than both the femoro-jugular (15.23 ± 7.00, n = 8) and the femoro-femoral cannulation configurations (13.49 ± 1.44, n = 8).
Full text links
Related Resources
Trending Papers
Haemodynamic monitoring during noncardiac surgery: past, present, and future.Journal of Clinical Monitoring and Computing 2024 April 31
2024 AHA/ACC/AMSSM/HRS/PACES/SCMR Guideline for the Management of Hypertrophic Cardiomyopathy: A Report of the American Heart Association/American College of Cardiology Joint Committee on Clinical Practice Guidelines.Circulation 2024 May 9
Obesity pharmacotherapy in older adults: a narrative review of evidence.International Journal of Obesity 2024 May 7
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app