We have located links that may give you full text access.
Health worker exposure risk during inhalation sedation with sevoflurane using the (AnaConDa®) anaesthetic conserving device.
INTRODUCTION AND OBJECTIVE: Occupational exposure to sevoflurane should not exceed 2 ppm. During inhalation sedation with sevoflurane using the anaesthetic conserving device (AnaConDa(®)) in the post-anaesthesia care unit, waste gases can be reduced by gas extraction systems or scavenging devices such as CONTRAfluran™. However, the efficacy of these methods has not been clearly established. To determine the safest scenario for healthcare workers during inhalation sedation with sevoflurane in the post-surgical intensive care unit.
MATERIALS AND METHODS: An experimental study on occupational exposure was conducted in a post-cardiothoracic care unit during March-August 2009. The measurements were performed in four post-cardiac surgery sedated adults in post-surgical intensive care unit and four nurses at the bedside, and at four points: scenario A, inhalation sedation without gas extraction system or contrafluran as a reference scenario; scenario B, applying a gas extraction system to the ventilator; scenario C, using contrafluran; and scenario 0, performing intravenous isolation sedation. Sevoflurane concentrations were measured in the nurses' breathing area during patient care, and at 1.5 and 8 m from the ventilator using diffusive passive monitor badges.
RESULTS: All badges corresponding to the nurses' breathing area were below 2 ppm. Levels of sevoflurane detected using prevention systems were lower than that in the control situation. Only one determination over 2 ppm was found, corresponding to the monitor placed nearest the gas outlet of the ventilator in scenario A. Trace concentrations of sevoflurane were found in scenario 0 during intravenous sedation.
CONCLUSIONS: Administration of sevoflurane through the AnaConDa(®) system during inhalation sedation in post-surgical intensive care units is safe for healthcare workers, but gas extraction systems or scavenging systems, such as CONTRAfluran™ should be used to reduce occupational exposure as much as possible.
MATERIALS AND METHODS: An experimental study on occupational exposure was conducted in a post-cardiothoracic care unit during March-August 2009. The measurements were performed in four post-cardiac surgery sedated adults in post-surgical intensive care unit and four nurses at the bedside, and at four points: scenario A, inhalation sedation without gas extraction system or contrafluran as a reference scenario; scenario B, applying a gas extraction system to the ventilator; scenario C, using contrafluran; and scenario 0, performing intravenous isolation sedation. Sevoflurane concentrations were measured in the nurses' breathing area during patient care, and at 1.5 and 8 m from the ventilator using diffusive passive monitor badges.
RESULTS: All badges corresponding to the nurses' breathing area were below 2 ppm. Levels of sevoflurane detected using prevention systems were lower than that in the control situation. Only one determination over 2 ppm was found, corresponding to the monitor placed nearest the gas outlet of the ventilator in scenario A. Trace concentrations of sevoflurane were found in scenario 0 during intravenous sedation.
CONCLUSIONS: Administration of sevoflurane through the AnaConDa(®) system during inhalation sedation in post-surgical intensive care units is safe for healthcare workers, but gas extraction systems or scavenging systems, such as CONTRAfluran™ should be used to reduce occupational exposure as much as possible.
Full text links
Related Resources
Trending Papers
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app