Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting.
OBJECTIVE: To investigate the precision and accuracy of continuous pulse contour cardiac output (PCCO) compared with intermittent transcardiopulmonary (TCPCO) and pulmonary artery thermodilution (TDCO) measurements in patients undergoing minimally invasive coronary bypass surgery (MIDCAB).
DESIGN: Prospective, controlled, clinical study.
SETTING: University hospital.
PARTICIPANTS: Twelve patients undergoing MIDCAB.
INTERVENTIONS: Thirty-six measurements of PCCO and thermodilution cardiac output (CO) were simultaneously performed after the start of surgery, during bypass grafting, and at the end of surgery. TCPCO and TDCO were simultaneously assessed by three injections of ice-cold saline randomly spread over the respiratory cycle. The pulse contour device was initially calibrated with an additional set of aortic thermodilution measurements.
MEASUREMENTS AND MAIN RESULTS: Absolute values of CO ranged between 1.6 and 9.2 L/min. A close agreement among the three techniques was observed at all measurements. Mean bias between PCCO and TDCO and TCPCO was 0.003 L/min (2 SD of differences between methods = 1.26 L/min) and 0.27 L/min (2 SD of differences between methods = 1.16 L/min), respectively. The correlation coefficients were r2 = 0.90 for TCPCO versus PCCO and r2 = 0.88 for TDCO versus PCCO.
CONCLUSION: The results of the present study show that compared with thermodilution CO, pulse contour analysis enables accurate measurement of continuous CO in patients undergoing MIDCAB.
DESIGN: Prospective, controlled, clinical study.
SETTING: University hospital.
PARTICIPANTS: Twelve patients undergoing MIDCAB.
INTERVENTIONS: Thirty-six measurements of PCCO and thermodilution cardiac output (CO) were simultaneously performed after the start of surgery, during bypass grafting, and at the end of surgery. TCPCO and TDCO were simultaneously assessed by three injections of ice-cold saline randomly spread over the respiratory cycle. The pulse contour device was initially calibrated with an additional set of aortic thermodilution measurements.
MEASUREMENTS AND MAIN RESULTS: Absolute values of CO ranged between 1.6 and 9.2 L/min. A close agreement among the three techniques was observed at all measurements. Mean bias between PCCO and TDCO and TCPCO was 0.003 L/min (2 SD of differences between methods = 1.26 L/min) and 0.27 L/min (2 SD of differences between methods = 1.16 L/min), respectively. The correlation coefficients were r2 = 0.90 for TCPCO versus PCCO and r2 = 0.88 for TDCO versus PCCO.
CONCLUSION: The results of the present study show that compared with thermodilution CO, pulse contour analysis enables accurate measurement of continuous CO in patients undergoing MIDCAB.
Full text links
Related Resources
Trending Papers
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app