Feasibility of a novel atrial mechanical sensing method for leadless atrioventricular synchronous pacing.
Journal of Cardiovascular Electrophysiology 2024 March 28
INTRODUCTION: Few studies have demonstrated a high degree of atrioventricular (AV) synchrony with a new leadless pacemaker called Micra AV. Our group previously reported a new and unique sensing method. We believe that this novel sensing method, "Simplified A3 method," might facilitate better AV synchrony than a conventional sensing method in almost all cases with Micra AV implantation.
METHODS: We conducted a non-randomized retrospective study comparing the two pacing methods at two centers. From December 1, 2021 to October 31, 2022, Micra AV was implanted for 32 patients at the two centers. Twenty of the 32 patients with sinus rhythm and complete AV block were included in this study. In Group 1, the conventional setting was programmed as follows: auto A3 threshold, auto A3 window-end (WE), and auto A4 threshold turned on during hospitalization. In Group 2, the "Simplified A3 method" was programmed as follows: auto A3 threshold and auto A3WE turned off. Instead, an intentionally prolonged A3WE (850-1000 ms) and low A3 threshold (A3 signal amplitude + 0.5-1.0 m/s2 ) were programmed.
RESULTS: Twenty patients were analyzed. In Group 2, AV synchrony (%AMVp) was significantly higher at the first outpatient clinic (63.0 ± 5.7% vs 81.0 ± 4.2%, p = .03). A3 threshold was significantly lower in Group 2 (5.9 ± 0.7 m/s2 , p < .05 vs 2.3 ± 0.5 m/s2 , p < .05).
CONCLUSION: Our novel sensing method might be a more feasible sensing method for obtaining higher AV synchrony than the conventional algorithm.
METHODS: We conducted a non-randomized retrospective study comparing the two pacing methods at two centers. From December 1, 2021 to October 31, 2022, Micra AV was implanted for 32 patients at the two centers. Twenty of the 32 patients with sinus rhythm and complete AV block were included in this study. In Group 1, the conventional setting was programmed as follows: auto A3 threshold, auto A3 window-end (WE), and auto A4 threshold turned on during hospitalization. In Group 2, the "Simplified A3 method" was programmed as follows: auto A3 threshold and auto A3WE turned off. Instead, an intentionally prolonged A3WE (850-1000 ms) and low A3 threshold (A3 signal amplitude + 0.5-1.0 m/s2 ) were programmed.
RESULTS: Twenty patients were analyzed. In Group 2, AV synchrony (%AMVp) was significantly higher at the first outpatient clinic (63.0 ± 5.7% vs 81.0 ± 4.2%, p = .03). A3 threshold was significantly lower in Group 2 (5.9 ± 0.7 m/s2 , p < .05 vs 2.3 ± 0.5 m/s2 , p < .05).
CONCLUSION: Our novel sensing method might be a more feasible sensing method for obtaining higher AV synchrony than the conventional algorithm.
Full text links
Related Resources
Trending Papers
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app