Accuracy of Noninvasive Blood Pressure Monitoring in Critically Ill Adults.
Journal of Intensive Care Medicine 2024 January 13
BACKGROUND: Blood pressure (BP) is routinely invasively monitored by an arterial catheter in the intensive care unit (ICU). However, the available data comparing the accuracy of noninvasive methods to arterial catheters for measuring BP in the ICU are limited by small numbers and diverse methodologies.
PURPOSE: To determine agreement between invasive arterial blood pressure monitoring (IABP) and noninvasive blood pressure (NIBP) in critically ill patients.
METHODS: This was a single center, observational study of critical ill adults in a tertiary care facility evaluating agreement (≤10% difference) between simultaneously measured IABP and NIBP. We measured clinical features at time of BP measurement inclusive of patient demographics, laboratory data, severity of illness, specific interventions (mechanical ventilation and dialysis), and vasopressor dose to identify particular clinical scenarios in which measurement agreement is more or less likely.
RESULTS: Of the 1852 critically ill adults with simultaneous IABP and NIBP readings, there was a median difference of 6 mm Hg in mean arterial pressure (MAP), interquartile range (1-12), P < .01. A logistic regression analysis identified 5 independent predictors of measurement discrepancy: increasing doses of norepinephrine (adjusted odds ratio [aOR] 1.10 [95% confidence interval, CI 1.08-1.12] P = .03 for every change in 5 µg/min), lower MAP value (aOR 0.98 [0.98-0.99] P < .01 for every change in 1 mm Hg), higher body mass index (aOR 1.04 [1.01-1.09] P = .01 for an increase in 1), increased patient age (aOR 1.31 [1.30-1.37] P < .01 for every 10 years), and radial arterial line location (aOR 1.74 [1.16-2.47] P = .04).
CONCLUSIONS: There was broad agreement between IABP and NIBP in critically ill patients over a range of BPs and severity of illness. Several variables are associated with measurement discrepancy; however, their predictive capacity is modest. This may guide future study into which patients may specifically benefit from an arterial catheter.
PURPOSE: To determine agreement between invasive arterial blood pressure monitoring (IABP) and noninvasive blood pressure (NIBP) in critically ill patients.
METHODS: This was a single center, observational study of critical ill adults in a tertiary care facility evaluating agreement (≤10% difference) between simultaneously measured IABP and NIBP. We measured clinical features at time of BP measurement inclusive of patient demographics, laboratory data, severity of illness, specific interventions (mechanical ventilation and dialysis), and vasopressor dose to identify particular clinical scenarios in which measurement agreement is more or less likely.
RESULTS: Of the 1852 critically ill adults with simultaneous IABP and NIBP readings, there was a median difference of 6 mm Hg in mean arterial pressure (MAP), interquartile range (1-12), P < .01. A logistic regression analysis identified 5 independent predictors of measurement discrepancy: increasing doses of norepinephrine (adjusted odds ratio [aOR] 1.10 [95% confidence interval, CI 1.08-1.12] P = .03 for every change in 5 µg/min), lower MAP value (aOR 0.98 [0.98-0.99] P < .01 for every change in 1 mm Hg), higher body mass index (aOR 1.04 [1.01-1.09] P = .01 for an increase in 1), increased patient age (aOR 1.31 [1.30-1.37] P < .01 for every 10 years), and radial arterial line location (aOR 1.74 [1.16-2.47] P = .04).
CONCLUSIONS: There was broad agreement between IABP and NIBP in critically ill patients over a range of BPs and severity of illness. Several variables are associated with measurement discrepancy; however, their predictive capacity is modest. This may guide future study into which patients may specifically benefit from an arterial catheter.
Full text links
Related Resources
Trending Papers
Obesity pharmacotherapy in older adults: a narrative review of evidence.International Journal of Obesity 2024 May 7
Haemodynamic monitoring during noncardiac surgery: past, present, and future.Journal of Clinical Monitoring and Computing 2024 April 31
SGLT2 Inhibitors in Kidney Diseases-A Narrative Review.International Journal of Molecular Sciences 2024 May 2
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app