Comparative performance of stimpod electromyography with mechanomyography for quantitative neuromuscular blockade monitoring.
Journal of Clinical Monitoring and Computing 2023 October 14
PURPOSE: The purpose of this study was to compare the Stimpod electromyograph neuromuscular blockade monitor to mechanomyography, which is widely considered to be the reference standard.
METHODS: The Stimpod electromyograph was used with its designated electrode array on the same hand as the mechanomyograph. Pairs of train-of-four measurements were recorded every 0.5-2 min. When the train-of-four count was zero on the electromyograph monitor, pairs of post tetanic count measurements were recorded every 2.5 min, instead of train-of-four measurements. Measurements were recorded from immediately after induction of anesthesia until just before emergence. Stimulation current was set to 60 mA with a duration of 200 microsec. The mechanomyography recording system recorded each twitch waveform for analysis. High resolution electromyograph waveforms were also recorded using a datalogger accessory provided by the manufacturer, facilitating inspection of individual waveforms. The administration of neuromuscular blocking drugs was left up to the discretion of the anesthesia care team.
RESULTS: Twenty-three patients contributed 1,088 data pairs suitable for analysis. Bland-Altman analysis of 415 pairs of train-of-four ratios showed a bias of 0.028 and limits of agreement of -0.18 and 0.24. Two hundred seventy-three train-of-four count data pairs were compared by Cohen's quadratically weighted kappa which was calculated to be 0.44, indicating moderate agreement. Three hundred thirty-eight post tetanic count data pairs were compared by Cohen's quadradically weighted kappa which was calculated to be 0.80, indicating substantial agreement.
CONCLUSION: The electromyograph produced results that were comparable to the mechanomyograph.
METHODS: The Stimpod electromyograph was used with its designated electrode array on the same hand as the mechanomyograph. Pairs of train-of-four measurements were recorded every 0.5-2 min. When the train-of-four count was zero on the electromyograph monitor, pairs of post tetanic count measurements were recorded every 2.5 min, instead of train-of-four measurements. Measurements were recorded from immediately after induction of anesthesia until just before emergence. Stimulation current was set to 60 mA with a duration of 200 microsec. The mechanomyography recording system recorded each twitch waveform for analysis. High resolution electromyograph waveforms were also recorded using a datalogger accessory provided by the manufacturer, facilitating inspection of individual waveforms. The administration of neuromuscular blocking drugs was left up to the discretion of the anesthesia care team.
RESULTS: Twenty-three patients contributed 1,088 data pairs suitable for analysis. Bland-Altman analysis of 415 pairs of train-of-four ratios showed a bias of 0.028 and limits of agreement of -0.18 and 0.24. Two hundred seventy-three train-of-four count data pairs were compared by Cohen's quadratically weighted kappa which was calculated to be 0.44, indicating moderate agreement. Three hundred thirty-eight post tetanic count data pairs were compared by Cohen's quadradically weighted kappa which was calculated to be 0.80, indicating substantial agreement.
CONCLUSION: The electromyograph produced results that were comparable to the mechanomyograph.
Full text links
Related Resources
Trending Papers
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Finerenone: From the Mechanism of Action to Clinical Use in Kidney Disease.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app