Comparision of cardiac output measured by transthoracic echocardiography with continuous cardiac output measured by pressure recording analytical method.
BACKGROUND: Low cardiac output is a common complication following cardiac surgery and it is associated with higher mortality in the pediatric population. A gold standard method for cardiac output (CO) monitoring in the pediatric population is lacking. The present study was conducted to validate cardiac output and cardiac index measured by transthoracic echocardiography and Pressure recording analytical method, a continuous pulse contour method, MostCareUp in postoperative pediatric cardiac surgical patients.
MATERIALS AND METHODS: This was a prospective observational clinical study conducted at a tertiary care hospital. A total of 23 pediatric patients weighed between 2 and 20 kg who had undergone elective cardiac surgery were included in the study.
RESULTS: Spearman's correlation coefficient of CO between transthoracic echocardiography (TTE) and Pressure Recording Analytical Method (PRAM) showed of positive correlation (r = 0.69, 95% Confidence interval 0.59-0.77, P < 0.0001) Linear regression equations for CO between TTE and PRAM were y = 0.55 + 0.88x (R2 = 0.46, P < 0.0001). (y = PRAM, x = TTE), respectively. Bland- Altman plot for CO between TTE and PRAM showed a bias of -0.397 with limits of the agreement being -2.01 to 1.22. Polar plot analysis showed an angular bias of 6.55° with radial limits of the agreement being -21.46 to 34.58 for CO and angular bias of 6.22° with radial limits of the agreement being -22.4 to 34.84 for CI.
CONCLUSION: PRAM has shown good trending ability for cardiac output. However, values measured by PRAM are not interchangeable with the values measured by transthoracic echocardiography.
MATERIALS AND METHODS: This was a prospective observational clinical study conducted at a tertiary care hospital. A total of 23 pediatric patients weighed between 2 and 20 kg who had undergone elective cardiac surgery were included in the study.
RESULTS: Spearman's correlation coefficient of CO between transthoracic echocardiography (TTE) and Pressure Recording Analytical Method (PRAM) showed of positive correlation (r = 0.69, 95% Confidence interval 0.59-0.77, P < 0.0001) Linear regression equations for CO between TTE and PRAM were y = 0.55 + 0.88x (R2 = 0.46, P < 0.0001). (y = PRAM, x = TTE), respectively. Bland- Altman plot for CO between TTE and PRAM showed a bias of -0.397 with limits of the agreement being -2.01 to 1.22. Polar plot analysis showed an angular bias of 6.55° with radial limits of the agreement being -21.46 to 34.58 for CO and angular bias of 6.22° with radial limits of the agreement being -22.4 to 34.84 for CI.
CONCLUSION: PRAM has shown good trending ability for cardiac output. However, values measured by PRAM are not interchangeable with the values measured by transthoracic echocardiography.
Full text links
Related Resources
Trending Papers
Haemodynamic monitoring during noncardiac surgery: past, present, and future.Journal of Clinical Monitoring and Computing 2024 April 31
Obesity pharmacotherapy in older adults: a narrative review of evidence.International Journal of Obesity 2024 May 7
2024 AHA/ACC/AMSSM/HRS/PACES/SCMR Guideline for the Management of Hypertrophic Cardiomyopathy: A Report of the American Heart Association/American College of Cardiology Joint Committee on Clinical Practice Guidelines.Circulation 2024 May 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app