Add like
Add dislike
Add to saved papers

Five-Year Outcomes of Endovascular Aortic Repair With the TREO Abdominal Endograft.

PURPOSE: Newer generation abdominal endografts, including Treo (Terumo Aortic, Sunrise, Florida), have shown optimal safety and effectiveness in treating abdominal aortic aneurysms (AAAs), even with hostile anatomy over the short- and mid-term. The durability of such results, however, is still a controversial issue, due to the paucity of long-term data. Our aim is to show the long-term outcomes of endovascular aortic repair of both standard and hostile AAAs with the Treo endograft on a cohort of patients treated between 2016 and 2017.

METHODS: We analyzed the postoperative follow-up of 37 consecutive patients who have undergone endovascular aortic repair (EVAR) with the Treo Endograft between 2016 and 2017, whose baseline clinical conditions, operative data, and short-term outcomes had been published in 2018. All patients were followed up by computed tomography angiography (CTA) at 6 and 12 months and 5 years postoperatively. Primary endpoints were aortic-related mortality, type I-III endoleak (EL), and reintervention rate. Secondary endpoints were the rates of type II ELs and aneurysm sac regression.

RESULTS: Of 37 patients, 27 had at least one criterion of anatomic hostility and 11 were performed outside the device-specific instructions for use (IFU). In the perioperative period, we observed 100% technical success, with no perioperative mortality. Over a mean follow-up of 5.5 years (66 months), 3 patients (8.1%) were lost to follow-up and 3 (8.1%) died of non-aortic causes (overall survival: 91.9%). One type IA EL of an AAA with a hostile neck (but within the IFU) and a type III EL of an AAA with standard anatomy were observed and treated by endovascular relining (overall reintervention rate: 5.5%). Four type II ELs were associated with aneurysm sac stability over time and are still under surveillance. Mean aneurysm shrinkage was 11.25±8.30 mm.

CONCLUSION: The optimal results of the Treo Endograft in terms of complication and reintervention rates reported over the mid-term by the current literature (ITA-ENDOBOOT registry) are maintained on the long term, both in case of hostile and friendly aortic anatomy, with a satisfactory shrinkage rate of the aneurysm sac.

CLINICAL IMPACT: The innovative characteristics of Treo and its short-term results are well-known and reported. The present case series contributes to the scientific validation of a new-generation abdominal aortic endograft over the long-term, focusing especially on its performance in treating AAAs with hostile anatomy. Its 5-years outcomes confirm the optimal results already reported over the short- and mid-term.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app