We have located links that may give you full text access.
Impact of Commissural Misalignment on Hydrodynamic Function Following Valve-in-Valve Intervention With the ACURATE neo.
JACC. Cardiovascular Interventions 2022 August 9
BACKGROUND: Limited evidence is available regarding valve-in-valve (VIV) intervention with the ACURATE neo transcatheter heart valve (THV). Low implantation has demonstrated leaflet interaction between the surgical bioprosthesis and the THV, leading to impaired hydrodynamic performance. It is unknown if commissural alignment (CA) can affect this phenomenon. Novel techniques have now been developed to achieve CA with the ACURATE neo THV.
OBJECTIVES: The aim of this study was to assess the impact of commissural misalignment (CMA) on hydrodynamic function following VIV intervention with the ACURATE neo THV using a bench model.
METHODS: VIV intervention was performed with the ACURATE neo (a self-expanding THV with supra-annular leaflet position) implanted deep in the surgical bioprosthetic aortic valve (Mitroflow). Hydrodynamic function at CA (0°) and 3 different degrees of CMA (30°, 60°, and 90°) was tested. As per the International Organization for Standardization, a regurgitant fraction <20% is considered optimal.
RESULTS: Following VIV, the central THV regurgitant fraction at 0°, 30°, 60°, and 90° of CMA was 8.6% ± 2.0%, 30.3% ± 12.0%, 42.6% ± 11.9%, and 66.7% ± 25.4% (P < 0.0001), respectively. On high-speed video there was no evidence of leaflet interaction at CA, whereas at 30°, 60°, and 90° of CMA there was clear evidence of THV leaflet interaction with those of the surgical valve, leading to impaired leaflet closure and to severe central THV regurgitation.
CONCLUSIONS: In VIV using the ACURATE neo THV at deep implantation, increasing degree of CMA was associated with THV leaflet interaction with those of the surgical valve and worsening regurgitant fraction. THV leaflet interaction was prevented when there was CA.
OBJECTIVES: The aim of this study was to assess the impact of commissural misalignment (CMA) on hydrodynamic function following VIV intervention with the ACURATE neo THV using a bench model.
METHODS: VIV intervention was performed with the ACURATE neo (a self-expanding THV with supra-annular leaflet position) implanted deep in the surgical bioprosthetic aortic valve (Mitroflow). Hydrodynamic function at CA (0°) and 3 different degrees of CMA (30°, 60°, and 90°) was tested. As per the International Organization for Standardization, a regurgitant fraction <20% is considered optimal.
RESULTS: Following VIV, the central THV regurgitant fraction at 0°, 30°, 60°, and 90° of CMA was 8.6% ± 2.0%, 30.3% ± 12.0%, 42.6% ± 11.9%, and 66.7% ± 25.4% (P < 0.0001), respectively. On high-speed video there was no evidence of leaflet interaction at CA, whereas at 30°, 60°, and 90° of CMA there was clear evidence of THV leaflet interaction with those of the surgical valve, leading to impaired leaflet closure and to severe central THV regurgitation.
CONCLUSIONS: In VIV using the ACURATE neo THV at deep implantation, increasing degree of CMA was associated with THV leaflet interaction with those of the surgical valve and worsening regurgitant fraction. THV leaflet interaction was prevented when there was CA.
Full text links
Related Resources
Trending Papers
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Finerenone: From the Mechanism of Action to Clinical Use in Kidney Disease.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app