Bioprosthetic valve fracture: Predictors of outcome and follow-up. Results from a multicenter study.
Catheterization and Cardiovascular Interventions 2021 October
OBJECTIVES: To evaluate outcome and its predictors of bioprosthetic valve fracture (BVF) in patients undergoing valve-in-valve transcatheter aortic valve replacement (VIV-TAVR).
BACKGROUND: BVF is feasible and reduces transvalvular gradients in VIV-TAVR-procedures, but follow-up-data and information on factors influencing the outcome are missing.
METHODS: The 81 cases of BVF-VIV-TAVR were collected from 14 international centers.
RESULTS: Predominantly transcatheter heart valve (THV) was implanted first, followed by BVF. VARC-2 defined device success was 93%, most failures were attributed to residual high gradients. Mean gradients decreased from 37 ± 13 mmHg to 10.8 ± 5.9 mmHg (p < 0.001). BVF reduced the gradient by 16 mmHg. During follow-up (FU, 281 ± 164 days) mean gradient remained stable (10.8 ± 5.9 mmHg at discharge, 12.4 ± 6.3 mmHg at FU, p = ns). In-hospital major adverse events occurred in 3.7%. Event-free survival at 276 ± 237.6 days was 95.4%. The linear mixed model identified balloon-expandable valves (BEV), Mitroflow surgical valve, stenotic surgical bioprostheses and balloon only 1 mm larger than the true internal diameter of the surgical valve as predictors for higher gradients.
CONCLUSIONS: BVF is safe and can significantly reduce gradients, which remain stable at FU. BEV, Mitroflow surgical valve, stenotic bioprostheses and balloon larger than the true internal diameter of the surgical valve of only 1 mm are predictors for higher final gradients.
BACKGROUND: BVF is feasible and reduces transvalvular gradients in VIV-TAVR-procedures, but follow-up-data and information on factors influencing the outcome are missing.
METHODS: The 81 cases of BVF-VIV-TAVR were collected from 14 international centers.
RESULTS: Predominantly transcatheter heart valve (THV) was implanted first, followed by BVF. VARC-2 defined device success was 93%, most failures were attributed to residual high gradients. Mean gradients decreased from 37 ± 13 mmHg to 10.8 ± 5.9 mmHg (p < 0.001). BVF reduced the gradient by 16 mmHg. During follow-up (FU, 281 ± 164 days) mean gradient remained stable (10.8 ± 5.9 mmHg at discharge, 12.4 ± 6.3 mmHg at FU, p = ns). In-hospital major adverse events occurred in 3.7%. Event-free survival at 276 ± 237.6 days was 95.4%. The linear mixed model identified balloon-expandable valves (BEV), Mitroflow surgical valve, stenotic surgical bioprostheses and balloon only 1 mm larger than the true internal diameter of the surgical valve as predictors for higher gradients.
CONCLUSIONS: BVF is safe and can significantly reduce gradients, which remain stable at FU. BEV, Mitroflow surgical valve, stenotic bioprostheses and balloon larger than the true internal diameter of the surgical valve of only 1 mm are predictors for higher final gradients.
Full text links
Related Resources
Trending Papers
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Finerenone: From the Mechanism of Action to Clinical Use in Kidney Disease.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app