Implications of hydrodynamic testing to guide sizing of self-expanding transcatheter heart valves for valve-in-valve procedures.
Catheterization and Cardiovascular Interventions 2020 September 2
AIMS: The commonly used valve-in-valve (VIV) app recommends sizing based on dimensions of both the transcatheter heart valve (THV) and bioprosthetic surgical valve. The implications of hydrodynamic testing to guide VIV sizing are poorly understood. This bench study assessed the hydrodynamic performance of different sizes of self-expanding supra-annular THVs in three different surgical aortic bioprostheses at different implantation depths.
METHODS: A small versus medium ACURATE neo (ACn), and a 26 mm versus 29 mm Evolut R were assessed after VIV implantation in 25 mm Mitroflow, Mosaic, and Magna Ease aortic surgical bioprostheses, at three implantation depths (+2 mm, -2 mm, and -6 mm).
RESULTS: The medium-sized ACn had lower gradients compared to the small ACn when the THV was implanted high (+2 mm, or -2 mm). The 29 mm Evolut R had lower gradients compared to a 26 mm Evolut R for all implantation depths, except for a depth of -2 mm in the 25 mm Mitroflow. The medium ACn and 29 mm Evolut R had larger effective orifice areas compared to the small ACn and 26 mm Evolut R, respectively. Both Evolut R sizes had acceptable regurgitant fractions (<15%), while both ACn sizes were above the acceptable performance criteria (>15%), at all implantation depths.
CONCLUSIONS: Use of a larger self-expanding THV was associated with superior hydrodynamic performance if the THV was implanted high. Hydrodynamic testing can provide additional information to the VIV app to help guide VIV sizing.
METHODS: A small versus medium ACURATE neo (ACn), and a 26 mm versus 29 mm Evolut R were assessed after VIV implantation in 25 mm Mitroflow, Mosaic, and Magna Ease aortic surgical bioprostheses, at three implantation depths (+2 mm, -2 mm, and -6 mm).
RESULTS: The medium-sized ACn had lower gradients compared to the small ACn when the THV was implanted high (+2 mm, or -2 mm). The 29 mm Evolut R had lower gradients compared to a 26 mm Evolut R for all implantation depths, except for a depth of -2 mm in the 25 mm Mitroflow. The medium ACn and 29 mm Evolut R had larger effective orifice areas compared to the small ACn and 26 mm Evolut R, respectively. Both Evolut R sizes had acceptable regurgitant fractions (<15%), while both ACn sizes were above the acceptable performance criteria (>15%), at all implantation depths.
CONCLUSIONS: Use of a larger self-expanding THV was associated with superior hydrodynamic performance if the THV was implanted high. Hydrodynamic testing can provide additional information to the VIV app to help guide VIV sizing.
Full text links
Related Resources
Trending Papers
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Finerenone: From the Mechanism of Action to Clinical Use in Kidney Disease.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app