AIM: Humidifying noninvasively ventilated preterm infants is critical to prevent dehydration of respiratory mucosa, but over-humidification can result in impaired airway patency and lung mechanics. This neonatal bench study investigated the humidity delivered using invasive and noninvasive humidification modes during nasal continuous positive airway pressure.
METHODS: The study was conducted at the neonatal intensive care unit of Emma Children's Hospital, the Netherlands, in March 2014. A mannequin was connected to a CareFusion Infant Flow SiPAP system, combined with a Fisher & Paykel MR850 humidifier and a Carefusion Infant Flow LP Generator. We measured the temperature, relative humidity and absolute humidity at the humidification chamber's expiratory port and at the patient's mask.
RESULTS: The absolute humidity at the mask was 35-40 mg H2 O/L at 38-39°C (relative humidity 74-80%) for the invasive mode of the humidifier and 23-27 mg H2 O/L at 34-35°C (relative humidity 63-70%) for the noninvasive mode. The absolute humidities exceeded the recommended values for the invasive mode of the humidifier, but not the noninvasive mode, and could be associated with increased condensation.
CONCLUSION: The absolute humidity delivered by the humidifier in invasive mode could be associated with increased condensation, which has been associated with airway impairment.
METHODS: The study was conducted at the neonatal intensive care unit of Emma Children's Hospital, the Netherlands, in March 2014. A mannequin was connected to a CareFusion Infant Flow SiPAP system, combined with a Fisher & Paykel MR850 humidifier and a Carefusion Infant Flow LP Generator. We measured the temperature, relative humidity and absolute humidity at the humidification chamber's expiratory port and at the patient's mask.
RESULTS: The absolute humidity at the mask was 35-40 mg H2 O/L at 38-39°C (relative humidity 74-80%) for the invasive mode of the humidifier and 23-27 mg H2 O/L at 34-35°C (relative humidity 63-70%) for the noninvasive mode. The absolute humidities exceeded the recommended values for the invasive mode of the humidifier, but not the noninvasive mode, and could be associated with increased condensation.
CONCLUSION: The absolute humidity delivered by the humidifier in invasive mode could be associated with increased condensation, which has been associated with airway impairment.
Full text links
Related Resources
Trending Papers
Systemic lupus erythematosus.Lancet 2024 April 18
Should renin-angiotensin system inhibitors be held prior to major surgery?British Journal of Anaesthesia 2024 May
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app