PURPOSE: Current solutions for navigated spine surgery remain hampered by restrictions in surgical workflow as well as a limited versatility and applicability. Against this background, we report the first experience of navigated spinal instrumentation with the mobile AIRO(®) intraoperative computed tomography (iCT) scanner.
METHODS: AIRO(®) iCT was used for navigated posterior spinal instrumentation of 170 screws in 23 consecutive patients operated on in our Department between the first use of the system in May 2014 and August 2014. The indications for AIRO(®) were based on the surgical region, anatomical complexity and the need for >3 segment instrumentation. Following navigated screw insertion, screw positions were confirmed intraoperatively by a second iCT scan. CT data on screw placement accuracy were retrospectively reviewed and analyzed by an independent observer.
RESULTS: AIRO(®)-based spinal navigation was easy to implement and successfully accomplished in all patients, adding around 18-34 min to the net surgery time. A systematic description of the authors' approach, setup in the OR and workflow integration of the AIRO(®) is presented. Analysis of screw placement accuracy revealed 9 (5.3%) screws with minor pedicle breaches (<2 mm). A total of 7 screws (4.1%) were misplaced >2 mm, resulting in an accuracy rate of 95.9%.
CONCLUSIONS: The AIRO(®) system is an easy-to-use and versatile iCT for navigated spinal instrumentation and provides high pedicle screw accuracy rates. Although the authors' experience suggests that the learning curve associated with AIRO(®)-based spinal navigation is steep, a systematic user-based approach to the technology is required.
METHODS: AIRO(®) iCT was used for navigated posterior spinal instrumentation of 170 screws in 23 consecutive patients operated on in our Department between the first use of the system in May 2014 and August 2014. The indications for AIRO(®) were based on the surgical region, anatomical complexity and the need for >3 segment instrumentation. Following navigated screw insertion, screw positions were confirmed intraoperatively by a second iCT scan. CT data on screw placement accuracy were retrospectively reviewed and analyzed by an independent observer.
RESULTS: AIRO(®)-based spinal navigation was easy to implement and successfully accomplished in all patients, adding around 18-34 min to the net surgery time. A systematic description of the authors' approach, setup in the OR and workflow integration of the AIRO(®) is presented. Analysis of screw placement accuracy revealed 9 (5.3%) screws with minor pedicle breaches (<2 mm). A total of 7 screws (4.1%) were misplaced >2 mm, resulting in an accuracy rate of 95.9%.
CONCLUSIONS: The AIRO(®) system is an easy-to-use and versatile iCT for navigated spinal instrumentation and provides high pedicle screw accuracy rates. Although the authors' experience suggests that the learning curve associated with AIRO(®)-based spinal navigation is steep, a systematic user-based approach to the technology is required.
Full text links
Related Resources
Trending Papers
Systemic lupus erythematosus.Lancet 2024 April 18
Should renin-angiotensin system inhibitors be held prior to major surgery?British Journal of Anaesthesia 2024 May
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app