Controlled Clinical Trial
Journal Article
Multicenter Study
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Intraoperative cone-beam computed tomography in oral and maxillofacial surgery using a C-arm prototype: first clinical experiences after treatment of zygomaticomaxillary complex fractures.

PURPOSE: To describe the first clinical applications of intraoperative cone-beam computed tomography with an integrated flat-panel detector in oral and maxillofacial surgery after surgical treatment of zygomaticomaxillary complex fractures

PATIENTS AND METHODS: Nine cone-beam computed tomography datasets of patients with zygomaticomaxillary complex fractures were intraoperatively acquired using a mobile isocentric C-arm (PowerMobil; Siemens Medical Solutions, Erlangen, Germany), including a flat-panel detector. Datasets based on 400, 200, and 100 fluoroscopic images were performed with different tube currents (4.6 mA, 3.3 mA, 2.3 mA, 1.2 mA, and 0.5 mA) and a current tube voltage of 100 kV. Postprocessing resulted in 15 different datasets available for comparison with corresponding preoperative computed tomography datasets. Four oral and maxillofacial surgeons and 2 experienced radiologists evaluated each dataset regarding noise, transition, and the delimitation of landmarks.

RESULTS: All examinations were successfully performed. Reconstructed datasets showed high-resolution images of all midfacial osseous structures in near-computed tomography quality. Regarding high-contrast structures, detailed analyses of datasets acquired in this study suggest that the parameters 400 projections, 1.2 mA, and 100 kV are sufficient. In terms of soft-tissue visualization, a higher level of mA seems preferable.

CONCLUSIONS: The tested prototype predicts a new era in cone-beam computed tomography imaging. The integration of a flat-panel detector will overcome the limitations of current available systems. The size of the field of view is increased allowing regularly the visualization of the whole facial skeleton. Particularly in cases of open reduction of unilateral fractures, the assessment of symmetry is of clinical value.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app