Hemodynamic Comparison between the Avalus and the Perimount Magna Ease Aortic Bioprosthesis up to 5 Years.
Thoracic and Cardiovascular Surgeon 2022 December 4
BACKGROUND: We aimed to compare hemodynamic performance of the Avalus (Medtronic) and the Perimount Magna Ease (PME, Edwards Lifesciences) bioprosthesis up to 5 years by serial echocardiographic examinations.
METHODS: In patients undergoing aortic valve replacement, 58 received PME prostheses between October 2007 and October 2008, and another 60 received Avalus prostheses between October 2014 and November 2015. To ensure similar baseline characteristics, we performed a propensity score matching based on left ventricular ejection fraction, age, body surface area, and aortic annulus diameter measured by intraoperative transesophageal echocardiography. Thereafter, 48 patients remained in each group. Mean age at operation was 67 ± 6 years and mean EuroSCORE-II was 1.7 ± 1.1. Both values did not differ significantly between the two groups.
RESULTS: At 1 year the mean pressure gradient (MPG) was 15.4 ± 4.3 mm Hg in the PME group and 14.7 ± 5.1 mm Hg in the Avalus group ( p = 0.32). The effective orifice area (EOA) was 1.65 ± 0.45 cm2 in the PME group and 1.62 ± 0.45 cm2 in the Avalus group ( p = 0.79). At 5 years the MPG was 16.6 ± 5.1 mm Hg in the PME group and 14.7 ± 7.1 mm Hg in the Avalus group ( p = 0.20). The EOA was 1.60 ± 0.49 cm2 in the PME group and 1.51 ± 0.40 cm2 in the Avalus group ( p = 0.38). Five-year survival was 88% in the PME group and 91% in the Avalus group ( p = 0.5). In the PME group, there were no reoperations on the aortic valve, whereas in the Avalus group three patients required a reoperation due to endocarditis.
CONCLUSION: Both bioprostheses exhibit similar hemodynamic performance during a 5-year follow-up.
METHODS: In patients undergoing aortic valve replacement, 58 received PME prostheses between October 2007 and October 2008, and another 60 received Avalus prostheses between October 2014 and November 2015. To ensure similar baseline characteristics, we performed a propensity score matching based on left ventricular ejection fraction, age, body surface area, and aortic annulus diameter measured by intraoperative transesophageal echocardiography. Thereafter, 48 patients remained in each group. Mean age at operation was 67 ± 6 years and mean EuroSCORE-II was 1.7 ± 1.1. Both values did not differ significantly between the two groups.
RESULTS: At 1 year the mean pressure gradient (MPG) was 15.4 ± 4.3 mm Hg in the PME group and 14.7 ± 5.1 mm Hg in the Avalus group ( p = 0.32). The effective orifice area (EOA) was 1.65 ± 0.45 cm2 in the PME group and 1.62 ± 0.45 cm2 in the Avalus group ( p = 0.79). At 5 years the MPG was 16.6 ± 5.1 mm Hg in the PME group and 14.7 ± 7.1 mm Hg in the Avalus group ( p = 0.20). The EOA was 1.60 ± 0.49 cm2 in the PME group and 1.51 ± 0.40 cm2 in the Avalus group ( p = 0.38). Five-year survival was 88% in the PME group and 91% in the Avalus group ( p = 0.5). In the PME group, there were no reoperations on the aortic valve, whereas in the Avalus group three patients required a reoperation due to endocarditis.
CONCLUSION: Both bioprostheses exhibit similar hemodynamic performance during a 5-year follow-up.
Full text links
Related Resources
Trending Papers
Obesity pharmacotherapy in older adults: a narrative review of evidence.International Journal of Obesity 2024 May 7
Haemodynamic monitoring during noncardiac surgery: past, present, and future.Journal of Clinical Monitoring and Computing 2024 April 31
SGLT2 Inhibitors in Kidney Diseases-A Narrative Review.International Journal of Molecular Sciences 2024 May 2
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app