Comparison of inspiratory effort with three variable-flow nasal continuous positive airway pressure devices in preterm infants: a cross-over study.

OBJECTIVE: Patient's work of breathing may vary between different neonatal nasal continuous positive airway pressure (NCPAP) devices. Therefore, we aimed to compare the inspiratory effort of three variable-flow NCPAP delivery systems used in preterm infants.

DESIGN: Cross-over study.

PATIENTS/SETTING: From June 2015 to August 2016, 20 preterm infants weighing ≤2500 g requiring NCPAP for mild respiratory distress syndrome were enrolled.

INTERVENTIONS: Each patient was successively supported by three randomly assigned variable-flow NCPAP systems (MedinCNO, Infant Flow and Servo-i) for 20 min while maintaining the same continuous positive airway pressure level as the patient was on before the study period.

MAIN OUTCOME MEASURES: Patients' inspiratory effort was estimated by calculating the sum of the difference between maximal inspiratory and baseline electrical activity of the diaphragm (∆EAdi) for 30 consecutive breaths, and after normalising this obtained value for the timing of the 30 breaths.

RESULTS: Physiological parameters (oxygen saturation measured by pulse oximetry, respiratory rate, heart beat, transcutaneous partial pressure CO2 ) and oxygen requirements remained identical between the three NCPAP systems. Although a wide variability in inspiratory effort could be observed, there were no statistically significant differences between the three systems for the sum of ∆EAdi for 30 breaths: CNO, 262 (±119) µV; IF, 352 (±262) µV; and SERVO-i, 286 (±126) µV, and the ∆EAdi reported on the timing of 30 breaths (sum ∆EAdi/s): CNO, 6.1 (±2.3) µV/s; IF, 7.9 (±4.9) µV/s; SERVO-i, 7.6 (±3.6) µV/s.

CONCLUSION: In a neonatal population of preterm infants, inspiratory effort is comparable between the three tested modern variable-flow NCPAP devices.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app