Journal Article
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Simultaneous Quantification of 13 Cannabinoids and Metabolites in Human Plasma by Liquid Chromatography Tandem Mass Spectrometry in Adult Epilepsy Patients.

BACKGROUND: A sensitive, robust method was developed and validated to quantitate 13 major natural cannabinoid parent and metabolite compounds in human plasma at or below 0.5 ng/mL.

METHODS: A liquid chromatography tandem mass spectrometry method was developed and validated to measure 13 cannabinoid compounds: cannabidiol (CBD), cannabidiolic acid, cannabidivarin, cannabinol, cannabigerol, cannabigerolic acid, cannabichromene, Δ-tetrahydocannabinol (THC), Δ-tetrahydrocannabinolic acid A (THCA), Δ-tetrahydrocannabivarin (THCV), 11-hydroxy-Δ-tetrahydrocannbinol (11-OH-THC), 11-nor-9-carboxy-Δ-tetrahydrocannbinol (THC-COOH), and 11-nor-9-carboxy-Δ-tetrahydrocannabinol glucuronide (THC-COOH-glu). Samples (200 µL) were extracted through protein precipitation and separated with a Kinetex EVO C18 column and a 65%-95% gradient of methanol and 0.2% ammonium hydroxide/H2O at a flow rate of 0.4 mL/min. Samples were obtained from patients with epilepsy receiving cannabis for the treatment of seizures.

RESULTS: The extracted lower limit of quantification was 0.05 ng/mL for CBD, cannabidivarin, cannabinol, and 11-OH-THC; 0.10 ng/mL for cannabidiolic acid, cannabigerol, cannabichromene, cannabigerolic acid, THC, THCA, and THCV; and 0.50 ng/mL for THC-COOH and THC-COOH-glu. Mean quality control intraday accuracy and precision for all analytes ranged 96.5%-104% and 2.7%-4.9%, respectively, whereas interday accuracy and precision ranged 98%-103.3% and 0.2%-3.6%, respectively. An absolute matrix effect was observed for some analytes, however, with minimal relative matrix effect. Lack of nonspecific drug binding to extraction glass and plasticware was verified. Patient CBD levels ranged from 0.135 to 11.13 ng/mL.

CONCLUSIONS: The validated method met FDA guidelines for bioanalytical assays precision and accuracy criteria. The assay reliably confirmed the use of particular medical cannabis formulations in patient samples as well as reliably measured low CBD concentrations from single-dose CBD exposure.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app