Use of a capillary specimen in the laboratory to verify a point-of-care international normalized ratio: Avoidance of a venipuncture in a pediatric setting.
International Journal of Laboratory Hematology 2019 Februrary
INTRODUCTION: Point-of-care (POC) international normalized ratio (INR) values above an institutional cutoff are confirmed in the laboratory using a gold standard venous specimen. This can be problematic in a pediatric setting.
METHOD: In this study, 449 consecutive POC INR results were compared to an INR performed in the laboratory on a capillary citrate specimen collected from the same finger-stick. The results were statistically analyzed.
RESULTS: The mean INR values from the CoaguChek XS and laboratory were 2.85 ± 1.19 and 2.63 ± 1.11, respectively. There was a good correlation between the methods with r = 0.97. Bland-Altman analysis indicated a bias of 0.22 favoring the CoaguChek XS, with 95% limits of agreement -0.29 to 0.72. Passing and Bablok method comparison resulted in a slope of 0.91 (y = 0.91x + 0.02). An INR of ≤0.5 was found between the methods in 89% of cases and 84% agreement was noted (κ = 0.69).
CONCLUSION: Comparing the capillary INR laboratory results to studies involving a venous specimen, the capillary specimen performed with equivalence. Thus, a capillary citrate specimen can be collected from the same finger-stick used to perform the POC INR for confirmation in the laboratory. This avoids the trauma of a venous collection in such a situation.
METHOD: In this study, 449 consecutive POC INR results were compared to an INR performed in the laboratory on a capillary citrate specimen collected from the same finger-stick. The results were statistically analyzed.
RESULTS: The mean INR values from the CoaguChek XS and laboratory were 2.85 ± 1.19 and 2.63 ± 1.11, respectively. There was a good correlation between the methods with r = 0.97. Bland-Altman analysis indicated a bias of 0.22 favoring the CoaguChek XS, with 95% limits of agreement -0.29 to 0.72. Passing and Bablok method comparison resulted in a slope of 0.91 (y = 0.91x + 0.02). An INR of ≤0.5 was found between the methods in 89% of cases and 84% agreement was noted (κ = 0.69).
CONCLUSION: Comparing the capillary INR laboratory results to studies involving a venous specimen, the capillary specimen performed with equivalence. Thus, a capillary citrate specimen can be collected from the same finger-stick used to perform the POC INR for confirmation in the laboratory. This avoids the trauma of a venous collection in such a situation.
Full text links
Related Resources
Trending Papers
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app