Patient-Specific Quality Assurance in Pencil Beam Scanning by 2-Dimensional Array.
PURPOSE: This study aimed to determine the characteristics of 2D ionization chamber array and the confidence limits of the gamma passing rate in pencil beam scanning proton therapy.
MATERIALS AND METHODS: The Varian ProBeam Compact spot-scanning system and the PTW OCTAVIUS 1500XDR array were used as a proton therapy system and detector, respectively. Our methods consisted of 2 parts: (1) the characteristics of the detector were tested and (2) patient-specific quality assurance was performed and evaluated by gamma analysis using dose-difference and distance-to-agreement criteria of 3% and 2 mm, respectively, with 123 treatment plans in head and neck, breast, chest, abdomen, and pelvic regions.
RESULTS: The PTW OCTAVIUS 1500XDR array had good reproducibility, uniformity, linearity, repetition rate, and monitor unit per spot within 0.1%, with accuracy, energy dependence, and measurement depth within 0.5%. The overall uncertainty of the PTW OCTAVIUS 1500XDR array was 2.49%. For field size and range shifter, using gamma analysis, the passing rate was 100%. The overall results of patient-specific quality assurance with the gamma evaluation were 98.9% ± 1.6% in 123 plans and confidence limit was 95.7%.
CONCLUSION: The PTW OTAVIUS 1500XDR offered effective performance in pencil beam scanning proton therapy.
MATERIALS AND METHODS: The Varian ProBeam Compact spot-scanning system and the PTW OCTAVIUS 1500XDR array were used as a proton therapy system and detector, respectively. Our methods consisted of 2 parts: (1) the characteristics of the detector were tested and (2) patient-specific quality assurance was performed and evaluated by gamma analysis using dose-difference and distance-to-agreement criteria of 3% and 2 mm, respectively, with 123 treatment plans in head and neck, breast, chest, abdomen, and pelvic regions.
RESULTS: The PTW OCTAVIUS 1500XDR array had good reproducibility, uniformity, linearity, repetition rate, and monitor unit per spot within 0.1%, with accuracy, energy dependence, and measurement depth within 0.5%. The overall uncertainty of the PTW OCTAVIUS 1500XDR array was 2.49%. For field size and range shifter, using gamma analysis, the passing rate was 100%. The overall results of patient-specific quality assurance with the gamma evaluation were 98.9% ± 1.6% in 123 plans and confidence limit was 95.7%.
CONCLUSION: The PTW OTAVIUS 1500XDR offered effective performance in pencil beam scanning proton therapy.
Full text links
Related Resources
Trending Papers
Review article: Recent advances in ascites and acute kidney injury management in cirrhosis.Alimentary Pharmacology & Therapeutics 2024 March 26
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app