Comparative Study
English Abstract
Journal Article
Randomized Controlled Trial
Add like
Add dislike
Add to saved papers

[Computer-assisted surgery for acetabular cup positioning in total hip arthroplasty: comparative prospective randomized study].

PURPOSE OF THE STUDY: Actetabular component malpositioning during total hip arthroplasty (THA) increases the risk of dislocation, reduces the range of motion, and can be the cause of early wear and loosening. There have been numerous reports on the optimal orientation of the acebaular component in THA. Lewinnek et al recommended an abduction angle of 40+/-10 degrees and an anteversion of 15+/-10 degrees for cup alignment in THA. In order to prevent malpostioned hip implants and improve the reproducibility of implant alignment in THA, numerous computer-assisted orthopedic systems have been described, using computed tomography (CT)-base or imageless navigation. Among the imageless systems available, one is based on Bone Morphing technology initially described by Stindel for computer-assisted knee arthroplasty and adapted for THA. The purpose of this study was to compare computer-assisted acetabular component insertion versus free hand placement.

MATERIAL AND METHODS: A controlled randomized matched prospective study was performed in two groups of 30 patients. The study was approved by the French Ethics Committee. In the first group, cup positioning was assisted by an imageless computer-assisted orthopedics system based on Bone Morphing(R) (CAOS+ group). In the control group, cup placement was free hand (CAOS- group). The same cementless cup was used in both groups. The same surgeon performed all procedures using an anterolateral approach. Cup anteversion and abduction angles were measured on 3D CT scan reconstructions obtained postoperatively for each patient by an independent observer using a special cup evaluation software.

RESULTS: There were 16 males and 14 females in each group, mean age was 62 years (range 24-80) years, and mean body mass index was 25 in each group. Mean additional time of the CAOS procedure was 12 minutes (range 8-20). Intraoperative subjective agreement of the surgeon with the computer guidance system demonstrated a high correlation in 23 cases, a weak correlation in six cases and poor correlation in one case. There were no statistical differences between the CAOS+ and the CAOS- group regarding means of the abduction and anteversion angles, but a significant range of variance, the lowest variations being observed in the CAOS+ group.

DISCUSSION: This study has shown the accuracy of cup positioning using a CT-free navigation system in a prospective randomized controlled protocol.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app