Novel high-resolution contrast agent ultrasound techniques HiFR CEUS and SR CEUS in combination with shear wave elastography, fat assessment and viscosity of liver parenchymal changes and tumors.

BACKGROUND: The continuous development of ultrasound techniques increasingly enables better description and visualization of unclear lesions. New ultrasound systems must be evaluated with regard to all these diagnostic possibilities.

METHODS: A multifrequency C1-7 convex probe (SC7-1M) with the new high-end system Resona A20 Series was used. Modern technologies, including HiFR CEUS, SR CEUS and multimodal tissue imaging with shear wave elastography (SWE), fat evaluation and viscosity measurements (M-Ref) were applied.

RESULTS: Of n = 70 (mean value 48,3 years±20,3 years, range 18-84 years) cases examined, a definitive diagnosis could be made in n = 67 cases, confirmed by reference imaging and/or follow-up. Of these, n = 22 cases were malignant changes (HCC (hepatocellular carcinoma) n = 9, CCC (cholangiocellular carcinoma) n = 3, metastases of colorectal carcinomas or recurrences of HCC n = 10). In all 12 cases of HCC or CCC, the elastography measurements using the shear wave technique (with values >2 m/s to 3.7 m/s) showed mean values of 2.3±0.31 m/s and a degree of fibrosis of F2 to F4. In n = 14 cases, changes in the fat measurement (range 0.51 to 0.72 dB/cm/MHz, mean values 0.58±0.12 dB/cm/MHz) in the sense of proportional fatty changes in the liver were detected. In the 4 cases of localized fat distribution disorders, the values were >0.7 dB/cm/MHz in the sense of significant fatty deposits in the remaining liver tissue. Relevant changes in the viscosity measurements with values >1.8 kPa were found in n = 31 cases, in n = 5 cases of cystic lesions with partially sclerosing cholangitis, in n = 13 cases of malignant lesions and in n = 9 cases post-interventionally, but also in n = 4 cases of benign foci with additional systemic inflammation.

CONCLUSIONS: The results are promising and show a new quality of ultrasound-based liver diagnostics. However, there is a need for further investigations with regard to the individual aspects, preferably on a multi-center basis.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app