Effect of varying X-ray tube voltage and additional filtration on image quality and patient dose in digital radiography system.

This study investigated the effect of varying x-ray tube voltage and additional filtration thicknesses on radiation dose and image quality in digital radiography system. The polymethylmethacrylate (PMMA) phantoms of different thicknesses simulating both the adult chest and abdomen and the pediatric patient's chest examinations were used. X-ray tube voltage range of 70-125 kVp was used for adult patient chest radiography, 70-100 kVp for adult patient abdominal radiography, and 50-70 kVp for pediatric 1-year-old chest examination. 0.1-0.3 mm Cu and 1.0 mm Al filters were used as additional filters. Patient doses were measured with an ionization chamber, considering the irradiation parameters recommended for radiographic examinations performed in radiology clinics in the EUR 16260 protocol. The Entrance Skin Dose (ESD) was calculated from the air kerma value measured at the entrance surface of the PMMA phantoms. Effective dose values were calculated by employing PCXMC 2.0 program. For image quality evaluations, CDRAD, LCD-4, Beam stop and Huttner test object used together with PMMA phantoms and Alderson RS-330 Lung/Chest phantom were used. Figure of Merit (FOM), which allows quantitative assessment in terms of image quality and patient dose, has been calculated. Based on the calculated FOM values were evaluated at the tube voltages and additional filter thicknesses recommended in the EUR 16260 protocol. Entrance Skin Dose and Inverse Image Quality Figure (IQFinv ) value obtained from contrast detail analysis decreased with increasing filter thickness and tube voltage. Decrease in ESD and IQFinv with increasing tube voltage without additional filter was 56% and 21% for adult chest radiography, 69% and 39% for adult abdominal radiography and 34% and 6% for 1-year-old pediatric chest radiography. When calculated FOM values are examined, it can be recommended to use a 0.1 mm Cu filter at 90 kVp and a 0.1 mm Cu + 1.0 mm Al filter at 125 kVp for adult chest radiography. For adult abdominal radiography, 0.2 mm Cu filter at 70 and 80 kVp and 0.1 mm Cu filter at 90 and 100 kVp were found to be appropriate. It was determined that the appropriate additional filter at 70 kVp for 1-year-old chest radiography was 1.0 mm Al+0.1 mm Cu.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app