Treatment of calcified lesions using a dedicated super-high pressure balloon. Multicenter optical coherence tomography registry.
BACKGROUND: Calcified lesions often lead to difficulty achieving optimal stent expansion. OPN non-compliant (NC) is a twin layer balloon with high rated burst pressure that may modify calcium effectively.
METHODS: Retrospective, multicenter registry in patients undergoing optical coherence tomography (OCT) guided intervention with OPN NC. Superficial calcification with > 180o arc and > 0.5 mm thickness, and/or nodular calcification with > 90o arc were included. OCT was performed in all cases before and after OPN NC, and after intervention. Primary efficacy endpoints were frequency of expansion (EXP) ≥80 % of the mean reference lumen area and mean final EXP by OCT, and secondary endpoints were calcium fractures (CF), and EXP ≥90 %.
RESULTS: 50 cases were included; 25 (50 %) superficial, and 25 (50 %) nodular. Calcium score of 4 in 42 (84 %) cases and 3 in 8 (16 %). OPN NC was used alone, or after other devices if further modification was needed, NC in 27 (54 %), cutting in 29 (58 %), scoring in 1 (2 %), IVL in 2 (4 %); or if non-crossable lesion, rotablation in 5 (10 %) cases. EXP ≥80 % was achieved in 40 (80 %) cases with mean final EXP post intervention of 85.7 % ± 8.9. CF were documented in 49 (98 %) cases; multiple in 37 (74 %). There were 1 flow limiting dissection requiring stent deployment and 3 non-cardiovascular related deaths in 6 months follow-up. No records of perforation, no-reflow or other major adverse events.
CONCLUSION: Among patients with heavy calcified lesions undergoing OCT guided intervention with OPN NC, acceptable expansion was achieved in most cases without procedure related complications.
METHODS: Retrospective, multicenter registry in patients undergoing optical coherence tomography (OCT) guided intervention with OPN NC. Superficial calcification with > 180o arc and > 0.5 mm thickness, and/or nodular calcification with > 90o arc were included. OCT was performed in all cases before and after OPN NC, and after intervention. Primary efficacy endpoints were frequency of expansion (EXP) ≥80 % of the mean reference lumen area and mean final EXP by OCT, and secondary endpoints were calcium fractures (CF), and EXP ≥90 %.
RESULTS: 50 cases were included; 25 (50 %) superficial, and 25 (50 %) nodular. Calcium score of 4 in 42 (84 %) cases and 3 in 8 (16 %). OPN NC was used alone, or after other devices if further modification was needed, NC in 27 (54 %), cutting in 29 (58 %), scoring in 1 (2 %), IVL in 2 (4 %); or if non-crossable lesion, rotablation in 5 (10 %) cases. EXP ≥80 % was achieved in 40 (80 %) cases with mean final EXP post intervention of 85.7 % ± 8.9. CF were documented in 49 (98 %) cases; multiple in 37 (74 %). There were 1 flow limiting dissection requiring stent deployment and 3 non-cardiovascular related deaths in 6 months follow-up. No records of perforation, no-reflow or other major adverse events.
CONCLUSION: Among patients with heavy calcified lesions undergoing OCT guided intervention with OPN NC, acceptable expansion was achieved in most cases without procedure related complications.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app