Extended Use of The Spanner® Temporary Prostatic Stent in Catheter-Dependent Men with Comorbidities.
Advances in Urology 2022
PURPOSE: This US FDA investigational device exemption (IDE) study evaluated the extended use of The Spanner® Temporary Prostatic Stent in catheter-dependent men with urinary retention who were not deemed candidates for corrective surgery but demonstrated bladder contractility.
MATERIALS AND METHODS: The Spanner was placed for 3 cycles of 30 days in catheter-dependent men with comorbid conditions, confirmed detrusor contractility, and catheter-associated discomfort. At each visit, postvoid residual, maximum flow rate, international prostate symptom score, quality of life, and adverse events were assessed. Voiding success was defined as PVR ≤ 150 ml at all visits.
RESULTS: One hundred seven men were enrolled at 8 US sites; 82/107 (76.6%) completed the trial, and 79/107 (73.8%) successfully maintained PVR ≤ 150 ml for the trial duration. Patients were 77.1 ± 10.6 years old; 63/107 (58.9%) were dependent on Foley and 40/107 (37.4%) on intermittent catheterization for 36.0 ± 39.3 days and 30.2 ± 45.8 days, respectively. 25/107 (23.4%) discontinuations were primarily due to voluntary patient withdrawal 9/107 (8.4%), investigator-initiated withdrawal 8/107 (7.5%), or lack of effectiveness 4/107 (3.7%). During Spanner use, the mean Q max was 11.2 ± 6.6, mean IPSS was 7.5 ± 6.4, and mean QOL was 2.0 ± 1.6. The most prevalent device-related adverse events were asymptomatic bacteriuria 25/107 (23.4%), discomfort 10/107 (9.4%), and urinary urgency 8/107 (7.5%). No device-related serious AEs were reported.
CONCLUSIONS: This study demonstrates that catheter-dependent men with sufficient bladder contractility can achieve volitional voiding and successful bladder drainage using The Spanner Temporary Prostatic Stent for extended periods of time.
MATERIALS AND METHODS: The Spanner was placed for 3 cycles of 30 days in catheter-dependent men with comorbid conditions, confirmed detrusor contractility, and catheter-associated discomfort. At each visit, postvoid residual, maximum flow rate, international prostate symptom score, quality of life, and adverse events were assessed. Voiding success was defined as PVR ≤ 150 ml at all visits.
RESULTS: One hundred seven men were enrolled at 8 US sites; 82/107 (76.6%) completed the trial, and 79/107 (73.8%) successfully maintained PVR ≤ 150 ml for the trial duration. Patients were 77.1 ± 10.6 years old; 63/107 (58.9%) were dependent on Foley and 40/107 (37.4%) on intermittent catheterization for 36.0 ± 39.3 days and 30.2 ± 45.8 days, respectively. 25/107 (23.4%) discontinuations were primarily due to voluntary patient withdrawal 9/107 (8.4%), investigator-initiated withdrawal 8/107 (7.5%), or lack of effectiveness 4/107 (3.7%). During Spanner use, the mean Q max was 11.2 ± 6.6, mean IPSS was 7.5 ± 6.4, and mean QOL was 2.0 ± 1.6. The most prevalent device-related adverse events were asymptomatic bacteriuria 25/107 (23.4%), discomfort 10/107 (9.4%), and urinary urgency 8/107 (7.5%). No device-related serious AEs were reported.
CONCLUSIONS: This study demonstrates that catheter-dependent men with sufficient bladder contractility can achieve volitional voiding and successful bladder drainage using The Spanner Temporary Prostatic Stent for extended periods of time.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app