Prospective, multicenter, randomized, controlled trial evaluating the performance of a novel combination powder vs hemostatic matrix in cardiothoracic operations.
Journal of Cardiac Surgery 2020 Februrary
AIM: This trial compared the hemostatic performance of a novel combination powder (CP) to a control hemostatic matrix (HM) in cardiothoracic operations.
METHODS: Patients meeting eligibility criteria were enrolled after providing informed consent. Subjects were randomized intraoperatively to receive CP (HEMOBLAST Bellows; Biom'up, France) or HM (FLOSEAL Hemostatic Matrix; Baxter Healthcare Corporation, Hayward, CA). Bleeding was assessed using a clinically validated, quantitative bleeding severity scale. The primary endpoint was total time to hemostasis (TTTH), from the start of device preparation, as an indicator of when a surgeon asks for a surgical hemostat until hemostasis was achieved. TTTH at 3 minutes was utilized for the primary analysis, while TTTH at 5 minutes was considered as a secondary endpoint.
RESULTS: A total of 105 subjects were enrolled across four institutions. The primary efficacy endpoint for the superiority of CP relative to HM for success at achieving hemostasis within 3 minutes was met, with 64.2% of the CP group achieving hemostasis compared with 9.6% of the HM group, a difference of 54.54% (37.4%-71.6%; P < .001 for superiority). The secondary efficacy endpoint was also met, with 92.5% of the CP group achieving hemostasis at 5 minutes versus 44.2% in the HM group, a difference of 48.2% (31.1%-65.4%; P < .001 for noninferiority). There were no device-related adverse events.
CONCLUSIONS: In this multicenter, randomized, controlled trial, comparison of CP to HM revealed CP superiority and noninferiority for TTTH at 3 and 5 minutes, respectively.
METHODS: Patients meeting eligibility criteria were enrolled after providing informed consent. Subjects were randomized intraoperatively to receive CP (HEMOBLAST Bellows; Biom'up, France) or HM (FLOSEAL Hemostatic Matrix; Baxter Healthcare Corporation, Hayward, CA). Bleeding was assessed using a clinically validated, quantitative bleeding severity scale. The primary endpoint was total time to hemostasis (TTTH), from the start of device preparation, as an indicator of when a surgeon asks for a surgical hemostat until hemostasis was achieved. TTTH at 3 minutes was utilized for the primary analysis, while TTTH at 5 minutes was considered as a secondary endpoint.
RESULTS: A total of 105 subjects were enrolled across four institutions. The primary efficacy endpoint for the superiority of CP relative to HM for success at achieving hemostasis within 3 minutes was met, with 64.2% of the CP group achieving hemostasis compared with 9.6% of the HM group, a difference of 54.54% (37.4%-71.6%; P < .001 for superiority). The secondary efficacy endpoint was also met, with 92.5% of the CP group achieving hemostasis at 5 minutes versus 44.2% in the HM group, a difference of 48.2% (31.1%-65.4%; P < .001 for noninferiority). There were no device-related adverse events.
CONCLUSIONS: In this multicenter, randomized, controlled trial, comparison of CP to HM revealed CP superiority and noninferiority for TTTH at 3 and 5 minutes, respectively.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app