We present the case of a 12-year-old girl with medically refractory epilepsy and a vagal nerve stimulator (VNS), who experienced severe obstructive sleep apnea (OSA) with respiratory events closely matching her VNS settings. We demonstrated a real-time decrease in OSA through an in-laboratory VNS titration study, decreasing her VNS frequency from 20 Hz to 10 Hz. We were able to demonstrate a baseline level of OSA by turning off the VNS. We then effectively treated her residual OSA with continuous positive airway pressure (CPAP). Novel to our case is that this in-laboratory VNS titration did not result in any subsequent increase in seizure frequency. After 5 months, her seizure frequency had decreased. Our case demonstrates that in-laboratory VNS titration can be an efficient tool for optimizing treatment of VNS-induced OSA and assert that polysomnography before VNS placement is important for guiding future care.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app