Evaluation of Acceptance Angle in Iodine-131 Single Photon Emission Computed Tomography Imaging with Monte Carlo Simulation.
Indian Journal of Nuclear Medicine : IJNM : the Official Journal of the Society of Nuclear Medicine, India 2019 January
Introduction: In iodine-131 (I-131) imaging, the image quality is degraded by scatter and penetration in a collimator. In this work, we assessed the penetrated and the scattered photon fractions in the photopeak energy window using Monte Carlo Simulation code.
Materials and Methods: The Siemens Medical System equipped with high-energy collimator was simulated. We evaluated the acceptance angle values on geometric, penetration, and scatter components in a separate file. Binary images in a data file are obtained and each one of them was imported in software ImageJ. Full-width at half-maximum (FWHM) and sensitivity were calculated and compared.
Results: The simulation data show that for the acceptance angle value equal to 4.845°, the geometric, scatter, and penetration components were 93.20%, 4.13%, and 2.67%, respectively. Moreover, the resolution is improved (FWHM = 7.21 mm, full width at tenth maximum = 12.36 mm) for a point source at 12 cm from the detector.
Conclusion: The small acceptance angle has a major impact on the image quality in I-131 single photon emission computed tomography.
Materials and Methods: The Siemens Medical System equipped with high-energy collimator was simulated. We evaluated the acceptance angle values on geometric, penetration, and scatter components in a separate file. Binary images in a data file are obtained and each one of them was imported in software ImageJ. Full-width at half-maximum (FWHM) and sensitivity were calculated and compared.
Results: The simulation data show that for the acceptance angle value equal to 4.845°, the geometric, scatter, and penetration components were 93.20%, 4.13%, and 2.67%, respectively. Moreover, the resolution is improved (FWHM = 7.21 mm, full width at tenth maximum = 12.36 mm) for a point source at 12 cm from the detector.
Conclusion: The small acceptance angle has a major impact on the image quality in I-131 single photon emission computed tomography.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app