Silicone oil-associated retinal light exposure under a surgical microscope.
Acta Ophthalmologica 2019 August
PURPOSE: Based on the hypothesis that silicone oil (SO) with a higher refractive index than water induces unexpected vignetting effects during surgeries, we aimed to investigate the relationship between the volume of SO and light intensity using a surgical microscope.
METHODS: Using a light-sensitive paper and model eye filled with varying volumes (0%, 50%, 62.5%, 75%, 87.5% and 100%) of SO, the light intensity of intraocular lens (IOL) with various refractive powers (0, 10 and 20 dioptres) illuminated by a surgical microscope was measured.
RESULTS: Light exposure density with 1.0% coaxial and oblique light was approximately 22-fold higher than that with 0.1% coaxial and oblique light. Further, it was approximately 34-fold higher in eyes with +20 D IOL than in those with no IOL. The density in eyes with 75% SO was the highest among all groups followed by that with 0% SO. Light exposure densities in the eyes with 75% and 0% SO were significantly higher than those with other volumes of SO. In SO-filled eyes, a microscope set with only an oblique light and a filter successfully reduced light exposure.
CONCLUSION: Silicone oil-related vision loss (SORVL) during SO removal surgeries might be due to increased light exposure on the macular retina caused by the SO-associated vignetting effect. SORVL could be prevented by placing a filter in the microscope during SO removal surgeries.
METHODS: Using a light-sensitive paper and model eye filled with varying volumes (0%, 50%, 62.5%, 75%, 87.5% and 100%) of SO, the light intensity of intraocular lens (IOL) with various refractive powers (0, 10 and 20 dioptres) illuminated by a surgical microscope was measured.
RESULTS: Light exposure density with 1.0% coaxial and oblique light was approximately 22-fold higher than that with 0.1% coaxial and oblique light. Further, it was approximately 34-fold higher in eyes with +20 D IOL than in those with no IOL. The density in eyes with 75% SO was the highest among all groups followed by that with 0% SO. Light exposure densities in the eyes with 75% and 0% SO were significantly higher than those with other volumes of SO. In SO-filled eyes, a microscope set with only an oblique light and a filter successfully reduced light exposure.
CONCLUSION: Silicone oil-related vision loss (SORVL) during SO removal surgeries might be due to increased light exposure on the macular retina caused by the SO-associated vignetting effect. SORVL could be prevented by placing a filter in the microscope during SO removal surgeries.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app