Comparative evaluation of CMAC and Truview picture capture device for endotracheal intubation in neonates and infants undergoing elective surgeries: A prospective randomized control trial.
Paediatric Anaesthesia 2018 December
BACKGROUND: Videolaryngoscopy has an established role in difficult airway management in adults. However, there is limited literature to support their efficacy in children. The Truview Picture Capture Device has shown promising results for endotracheal intubation in infants in the past. The CMAC videolaryngoscope has launched its novel infant Miller blade, but its performance has not been assessed clinically for routine intubation in infants and neonates. We hypothesized that the CMAC videolaryngoscope would reduce the total time to intubation as compared to the Truview Picture Capture Device in neonates and infants.
METHODS: After parental informed consent, 80 prospective infants posted for surgical procedures under general anesthesia were randomized to undergo intubation with either of the two. The two videolaryngoscopes were also compared in terms of time required for glottis view and intubation (primary outcome), modified Cormack and Lehane grade, first attempt and overall success rate, ease of intubation, number of attempts, and any complications.
RESULTS: The CMAC significantly reduced the time required for glottic view [8 s (5.25-9) vs 9 s (6.5-12); P = 0.02] and intubation [22 s (18-26) vs 26 s (21.5-32); P = 0.003]. The median difference (95% CI) for time to tracheal intubation and time to glottic view was 4 s (1-7) and 1 (0-4). It also improved the ease of intubation, the Cormack-Lehane grades, and first attempt success rate. Intubation with the CMAC was possible in 100% cases, whereas only 92.5% of patients could be intubated with the Truview. The failed intubations with the Truview could be successfully intubated with the CMAC.
CONCLUSION: The CMAC Miller blade reduced the total time taken for tracheal intubation and intubation difficulty as compared to the Truview Picture Capture Device and may be a better tool for intubation in infants.
METHODS: After parental informed consent, 80 prospective infants posted for surgical procedures under general anesthesia were randomized to undergo intubation with either of the two. The two videolaryngoscopes were also compared in terms of time required for glottis view and intubation (primary outcome), modified Cormack and Lehane grade, first attempt and overall success rate, ease of intubation, number of attempts, and any complications.
RESULTS: The CMAC significantly reduced the time required for glottic view [8 s (5.25-9) vs 9 s (6.5-12); P = 0.02] and intubation [22 s (18-26) vs 26 s (21.5-32); P = 0.003]. The median difference (95% CI) for time to tracheal intubation and time to glottic view was 4 s (1-7) and 1 (0-4). It also improved the ease of intubation, the Cormack-Lehane grades, and first attempt success rate. Intubation with the CMAC was possible in 100% cases, whereas only 92.5% of patients could be intubated with the Truview. The failed intubations with the Truview could be successfully intubated with the CMAC.
CONCLUSION: The CMAC Miller blade reduced the total time taken for tracheal intubation and intubation difficulty as compared to the Truview Picture Capture Device and may be a better tool for intubation in infants.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app